Accuracy of a new paired imaging technique for position correction in whole breast radiotherapy

نویسندگان

  • Saskia Petillion
  • Karolien Verhoeven
  • Caroline Weltens
  • Frank Van den Heuvel
چکیده

Image-guided position verification in breast radiotherapy is accurately performed with kilovoltage cone beam CT (kV-CBCT). The technique is, however, time-consuming and there is a risk for patient collision. Online position verification performed with orthogonal-angled mixed modality paired imaging is less time-consuming at the expense of inferior accuracy compared to kV-CBCT. We therefore investigated whether a new tangential-angled single modality paired imaging technique can reduce the residual error (RE) of orthogonal-angled mixed modality paired imaging. The latter was applied to 20 breast cancer patients. Tangential-angled single modality paired imaging was investigated in 20 breast and 20 breast cancer patients with locoregional lymph node irradiation. The central lung distance (CLD) residual error and the longitudinal residual error were determined during the first 5 treatment fractions. Off-line matching of the tangential breast field images, acquired after online position correction, was used. The mean, systematic, and random REs of each patient group were calculated. The systematic REs were checked for significant differences using the F-test. Tangential-angled single modality paired imaging significantly reduced the systematic CLD residual error of orthogonal-angled mixed modality paired imaging for the breast cancer patients, from 2.3 mm to 1.0 mm, and also significantly decreased the systematic longitudinal RE from 2.4 mm to 1.3 mm. PTV margins, which account for the residual error (PTVRE), were also calculated. The PTVRE margin needed to account for the RE of orthogonal-angled mixed modality paired imaging (i.e., 8 mm) was halved by tangential-angled single modality paired imaging. The differences between the systematic REs of tangential-angled single modality paired imaging of the breast cancer patients and the breast cancer patients with locoregional lymph node irradiation were not significant, yielding comparable PTVRE margins. In this study, we showed that tangential-angled single modality paired imaging is superior to orthogonal-angled mixed modality paired imaging to correct the position errors in whole breast radiotherapy.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Dosimetric comparison of whole breast radiotherapy using field in field and conformal radiotherapy techniques in early stage breast cancer

Background: We aimed to compare field-in-field technique (FIF) with conformal tangential field radiotherapy (CRT) in terms of dosimetric benefits for early stage breast cancer radiotherapy. Materials and Methods: Twenty consecutive left-side breast cancer patients who underwent breast-conserving surgery were included to the study. For each patient, two different treatment plans were cr...

متن کامل

Dosimetric comparison of 3-dimensional conformal and intensity-modulated radiotherapy techniques for whole breast irradiation in the prone and supine positions

Background: The aim of this study was to compare the differences of the dosimetric parameters between three-dimensional conformal radiotherapy (3D-CRT) and simultaneous-integrated boost intensity-modulated radiotherapy (SIB-IMRT) techniques in the prone and supine positions for breast irradiation. Materials and Methods: Ten patients underwent a computed tomography simulation in both the prone a...

متن کامل

Efficacy and workload analysis of a fixed vertical couch position technique and a fixed‐action–level protocol in whole‐breast radiotherapy

Quantification of the setup errors is vital to define appropriate setup margins preventing geographical misses. The no-action-level (NAL) correction protocol reduces the systematic setup errors and, hence, the setup margins. The manual entry of the setup corrections in the record-and-verify software, however, increases the susceptibility of the NAL protocol to human errors. Moreover, the impact...

متن کامل

Pretreatment quality control of single isocenter half- beam treatment planning technique using an amorphous silicon electronic portal-imaging device (EPID)

Introduction: Electronic portal imaging devices (EPIDs) are fundamentally used for instantaneous verification of the patient set‐up, block shape, and leaf positions during radiation therapy. In radiotherapy, situations arise in which an inclined PTV must be treated mutually with adjacent nodal regions. This methodology is most widely used for matching tangential/lateral breas...

متن کامل

Verification of the Accuracy of the Delivered Dose in Pelvic and Breast Cancer Radiotherapy by in-vivo Semi-Conductor Dosimetry

Introduction: Delivering maximum dose to tumor and minimum dose to normal tissues is the most important goal in radiotherapy. According to ICRU, the maximum acceptable uncertainty in the delivered dose compared to the prescribed dose should be lower than 5%, and this is because of the relationship between absorbed dose, tumor control and normal tissue damage. Absorbed dose accuracy is investiga...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 16  شماره 

صفحات  -

تاریخ انتشار 2015