[Contact lenses and intraocular lens power calculation: refractive surprise using the IOLMaster non-contact biometric method].

نویسندگان

  • M Díaz-Llopis
  • S García-Delpech
  • A Quijada
  • D Salom
  • P Udaondo
چکیده

Dear Sir, This letter is to alert about the danger of carrying out biometric calculations for the intraocular lens without being aware that the patient may be using a contact lens. This warning is valid even when the highest precision research technology is being utilised. At present, errors in the calculation of the intraocular lens (IOL) are the most frequent cause of lens exchanges as well as legal claims (1). The development of noncontact biometry with laser interferometry (IOL-MASTER, Zeiss) has increased the precision of said calculation, which is particularly important in transparent lens refractive surgery where the main objective is post-op emmetropy. We report the case of a 58-year old patient with budding lens opacity and 10/10 vision with -8.25 refraction in both eyes, who was sheduled for intraocular surgery and multifocal lens. In the period between the ophthalmological exploration and the measurements taken by the technical staff and the optician-optometrist, without informing the staff and while they were busy preparing the measurement, the patient placed his contact lenses on prior to the biometric readings. The biometric calculation was made with the Zeiss IOLMaster system (V.3.01; Carl Zeiss, Jena, Germany). The mean axial length and keratometry was in the RE of 26.21 mm/37.42 corneal diopters (D) and of 26.03 mm/38.63 D in the RE, which established a calculation for emmetropy of the multifocal intraocular lens (366D Acri.tec, constant A=118,3) with Haigis formula of +21 RE and +19 LE. Afte the RE surgery, the uncorrected visual acuity was below 1/10 with a value of –7,25. After repeating the biometric calculation in the contralateral eye (LE), the same result was obtained (+19,5). However, when we noticed that the patient was wearing a contact lens (-7D, radio 8,6, Purevision, Bausch&Lomb), we proceeded to withdraw it and repeat the calculation. The new results were: RE, 26.14 mm/44.77 D; LE: 26.03 mm/45.08 D. After exchanging the RE IOL and surgery with power adjusted IOL in LE, the final values were: RE, -0,25-0,25 at 20o, LE: 0,5 satisfactory bilateral emmetropy without need of eyeglasses. The ability of contact lenses, particularly hard and gas permeable lenses, to modify the biometric calculation of the intraocular lens by modifying the corneal curvature and therefore the corneal dioptries (K), sometimes for weeks after their withdrawal (2) is well known. However, after searching world literature in PubMed we found only one similar case related to a severe biometric error after carrying out biometric measurements while wearing disposable soft contact lenses (3). We conclude that even though we can utilize the latest technology for calculating intra-ocular lens values (IOLMASTER), it is very important to check that the patient it is not wearing contact lenses (including the soft and disposable type) in order to avoid surprises in refraction which could reach to 7 diopters, as we have seen in this case.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

IOLMaster keratometry for the calculation of toric intraocular lens power and orientation

In the implantation of intraocular lenses (IOL) stringent requirements are made on achieving the refractive target. This necessitates the exact determination of the basic data needed for IOL power calculation – both the axial length and, in particular, the corneal power – and additionally the cylinder power and its axis in the case of a toric lens. This gives rise to the question, especially fo...

متن کامل

Accuracy of axial length measurements from immersion B-scan ultrasonography in highly myopic eyes.

AIM To evaluate the accuracy of axial length (AL) measurements obtained from immersion B-scan ultrasonography (immersion B-scan) for intraocular lens (IOL) power calculation in patients with high myopia and cataracts. METHODS Immersion B-scan, contact A-scan ultrasonography (contact A-scan), and the IOLMaster were used to preoperatively measure the AL in 102 eyes from 102 patients who underwe...

متن کامل

Refractive outcomes of intraoperative wavefront aberrometry versus optical biometry alone for intraocular lens power calculation

PURPOSE To compare the outcomes of intraoperative wavefront aberrometry versus optical biometry alone for intraocular lens (IOL) power calculation in eyes undergoing cataract surgery with monofocal IOL implantation. METHODS Preoperative data were obtained with the IOLMaster. Intraoperative aphakic measurements and IOL power calculations were obtained in some patients with the optiwave refract...

متن کامل

مقایسه‌ی کراتومتری پنتاکم وکراتومتری IOL Master در محاسبه‌ی لنز داخل چشمی افراد با قرنیه‌ی طبیعی

Background and Objective: Proper method and machine for corneal evaluation is an important factor in many anterior segment interventions. This study was performed to compare the mean keratometry (K) readings obtained with an IOL Master and Pentacam for intraocular lens power calculation in normal subjects with no history of refractive surgery. Materials and Methods: Mean K values were obtained ...

متن کامل

Mastering the IOLMaster 16

WHILE the IOLMaster has become an indispensable tool for pre-operative assessment of cataract patients, it is important to understand its limitations, reported British researchers at the 8th ESCRS Winter Refractive Surgery meeting. “Accurate and precise biometry is one of the key factors in obtaining a good refractive outcome after cataract surgery.An error of only 1.0 mm in axial length will r...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Archivos de la Sociedad Espanola de Oftalmologia

دوره 83 1  شماره 

صفحات  -

تاریخ انتشار 2008