Comparison of Three Preamplifier Technologies: Variation of Input Impedance and Noise Figure With B0 Field Strength
نویسندگان
چکیده
Introduction Array coil preamplifiers are subjected to large magnetic fields since they are placed close to each coil to minimize SNR losses due to cabling. Noise figure (NF) and input impedance are key parameters that must vary minimally with magnetic field strength and orientation [1,2] to ensure there are no differences in performance between the bench and the bore. In this work we compare the sensitivity to magnetic fields of three amplifiers based on different semiconductor technologies. Methods The following commercially available devices (DUTs) were chosen for comparison (in order of decreasing electron mobility): MGA-53543 (Avago, USA) GaAs pHEMT RFIC; MAR-8A+ (Mini-Circuits, USA) InGaP HBT in Darlington configuration; and BF998 (various vendors) N-channel Si dual-gate MOSFET. An initial baseline noise figure measurement was performed in a radiofrequency (RF) lab in order to obtain a reference zero-field value. The measurement system (Fig. 1, [3]) was then placed outside the 5 gauss field line of a 9.4T magnet (205 mm bore, Magnex, UK). Double-shielded coaxial cables (RG223) 6 meters in length were used to connect the measurement system to the DUT through the magnet’s RF cage. Each amplifier was positioned in three orthogonal orientations within the magnet bore and scattering (S) parameters were measured at 130 MHz at nine B0 field strengths between 0 T and 9.4 T, varying the field as described in [1]. For each amplifier the orientation that produced the greatest change in S-
منابع مشابه
Variation of Preamplifier Noise Figure With B0 Field Strength
Introduction High-field arrays with large numbers of coils require preamplifiers located close to the coils to minimize SNR losses incurred in cables. The design of low-noise preamplifiers for high-field MRI must thus ensure there is no performance degradation due to the presence or orientation of high magnetic fields. Preamplifier scattering (S) parameters have been known to vary with B0 field...
متن کاملA self-built two stage stable preamplifier for small coils
Introduction: Coil arrays for signal reception are widely used in MRI to obtain high local SNR [1] and for improved imaging speed [2]. Increasing the channel number in the coil array, the single coil element in every channel becomes smaller, respectively. Thus, higher gain is required for preamplifiers to improve low signal level to acceptable values. However, high gain is a conflicting goal wi...
متن کاملNoise Contributions in Receive Coil Arrays
Introduction: Receive coil arrays are used in MR to optimize Signal to Noise Ration (SNR) of measured signal. The engineer’s goal is to build such arrays with high filling factor using electrical components with minimal losses, and decoupled coil elements such that elements noise and noise correlation are reduced. Theories for SNR maximization are always based on theoretical assumptions and a g...
متن کاملA New Ultra-Wideband Low Noise Amplifier With Continuous Gain Control
This paper presents a new variable gain low noise amplifier (VG-LNA) for ultra-wideband (UWB) applications. The proposed VG-LNA uses a common-source (CS) with a shunt-shunt active feedback as an input stage to realize input matching and partial noise cancelling. An output stage consists of a gain-boosted CS cascode and a gain control circuit that moves the high resonant frequency to higher freq...
متن کاملA Low Input Impedance MRI Preamplifier Using a Purely Capacitive Feedback Network
R. Oppelt, M. Vester Corporate Technology, Siemens AG, Erlangen, Germany, Medical Solutions, Siemens AG, Erlangen, Germany Introduction: MRI preamplifiers need to meet a number of technical requirements that are not easily to be achieved simultaneously. Major challenges are a low noise figure (e.g. <= 0.5 dB), a stable gain and a high dynamic range. In addition, the small signal input impedance...
متن کامل