OBJECTIVES To compare two hematological analyzers-the DxH-800 (DxH; Beckman-Coulter, Miami, FL) and XN-2000 (XN; Sysmex, Kobe, Japan)-with the Cell-Dyn Sapphire (SAPH; Abbott, Santa Clara, CA). METHODS We analyzed 4,375 samples. Slide reviews were made in the presence of blast, abnormal lymphocyte, and immature granulocyte (IG) flags or nucleated RBC (NRBC) count. RESULTS The analyzers exhibited excellent correlations for CBC and neutrophils but displayed a limit correlation for lymphocytes. The XN did not miss circulating blasts (0.5%-95% in microscopy). For NRBCs, the XN demonstrated a sensitivity of 90%; DxH, 74%; and SAPH, 29%. Only the XN demonstrated a correlation with microscopy, permitting a WBC six-part differential until 15% of NRBCs. The XN and DxH gave useful IG counts with a cutoff less than 5% and a WBC level more than 2,500/mm(3). For abnormal lymphocytes detection, only XN demonstrated sensitivity of more than 95%, but its specificity of 54% requires adaptation. CONCLUSION The XN increases the sensitivity of abnormal cell detection compared with the other counters, permitting a seven-part differential between predefined levels, decreasing the slide review from 20% to 9%.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of the Beckman Coulter UniCel DxH 800 and Abbott Diagnostics Cell-Dyn Sapphire hematology analyzers on pediatric and neonatal specimens in a tertiary care hospital.

We evaluated the new UniCel DxH 800 hematology analyzer (Beckman Coulter, Miami, FL) vs the Cell-Dyn Sapphire (Abbott Diagnostics, Santa Clara, CA) using 156 pediatric specimens in Microtainer tubes (Becton Dickinson, Franklin Lakes, NJ). The CBC and differential showed good interinstrument correlation, including WBCs (r = 0.995), RBCs (r = 0.992), hemoglobin (r = 0.998), mean corpuscular volum...

متن کامل

Comparison of automated differential blood cell counts from Abbott Sapphire, Siemens Advia 120, Beckman Coulter DxH 800, and Sysmex XE-2100 in normal and pathologic samples.

Reliable automated blood cell characterization and quantification remain challenging in pathologic samples, whereas slide reviews due to unnecessary flagging should be avoided. We compared 4 modern hematology analyzers-Abbott Sapphire, Siemens Advia 120, Sysmex XE-2100, and Beckman Coulter DxH 800-regarding complete blood cell count (CBC), leukocyte differential count, and flagging efficacy in ...

متن کامل

Evaluation of the Beckman Coulter UniCel DxH 800, Beckman Coulter LH 780, and Abbott Diagnostics Cell-Dyn Sapphire hematology analyzers on adult specimens in a tertiary care hospital.

We evaluated the new Beckman Coulter DxH 800 hematology analyzer (Beckman Coulter, Miami, FL) vs the Abbott Diagnostics Cell-Dyn Sapphire (Abbott Diagnostics, Santa Clara, CA) and Beckman Coulter LH 780 hematology analyzers using 430 adult specimens. The DxH 800 provided a CBC and differential that correlated well with those of the Sapphire and LH 780, with most parameters showing correlation c...

متن کامل

An Overview on: The Red Blood Cell Histogram

The RBC histogram is an integral part of automated hematology analysis and is now routinely available on all automated cell counters. This histogram and other associated complete blood count (CBC) parameters have been found abnormal in various hematological conditions and may provide major clues in the diagnosis and management of significant red cell disorders. In addition, it is frequently use...

متن کامل

Initial performance evaluation of the UniCel® DxH 800 Coulter® cellular analysis system

The Beckman Coulter UniCel® DxH 800 is a hematology analyzer incorporating new electronic and mechanical design with advanced algorithm technology to perform CBC, white blood cell (WBC) differential, nucleated red blood cell (NRBC), and reticulocyte analysis. Evaluation of this instrument was performed in our 800-bed tertiary care hospital and specifically centered upon the correlation of WBC, ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • American journal of clinical pathology

دوره 140 6  شماره 

صفحات  -

تاریخ انتشار 2013