Quality assurance of a helical tomotherapy machine.
نویسندگان
چکیده
Helical tomotherapy has been developed at the University of Wisconsin, and 'Hi-Art II' clinical machines are now commercially manufactured. At the core of each machine lies a ring-gantry-mounted short linear accelerator which generates x-rays that are collimated into a fan beam of intensity-modulated radiation by a binary multileaf, the modulation being variable with gantry angle. Patients are treated lying on a couch which is translated continuously through the bore of the machine as the gantry rotates. Highly conformal dose-distributions can be delivered using this technique, which is the therapy equivalent of spiral computed tomography. The approach requires synchrony of gantry rotation, couch translation, accelerator pulsing and the opening and closing of the leaves of the binary multileaf collimator used to modulate the radiation beam. In the course of clinically implementing helical tomotherapy, we have developed a quality assurance (QA) system for our machine. The system is analogous to that recommended for conventional clinical linear accelerator QA by AAPM Task Group 40 but contains some novel components, reflecting differences between the Hi-Art devices and conventional clinical accelerators. Here the design and dosimetric characteristics of Hi-Art machines are summarized and the QA system is set out along with experimental details of its implementation. Connections between this machine-based QA work, pre-treatment patient-specific delivery QA and fraction-by-fraction dose verification are discussed.
منابع مشابه
QA for helical tomotherapy: report of the AAPM Task Group 148.
Helical tomotherapy is a relatively new modality with integrated treatment planning and delivery hardware for radiation therapy treatments. In view of the uniqueness of the hardware design of the helical tomotherapy unit and its implications in routine quality assurance, the Therapy Physics Committee of the American Association of Physicists in Medicine commissioned Task Group 148 to review thi...
متن کاملEnhanced efficiency in helical tomotherapy quality assurance using a custom-designed water-equivalent phantom.
Tomotherapy is an image-guided, intensity-modulated radiation therapy system that delivers highly conformal dose distributions in a helical fashion. This system is also capable of acquiring megavoltage computed-tomography images and registering them to the planning kVCT images for accurate target localization. Quality assurance (QA) of this device is time intensive, but can be expedited by impr...
متن کاملQuality assurance of the multileaf collimator with helical tomotherapy: design and implementation.
Quality assurance (QA) of the multileaf collimator (MLC) is a critical step for the delivery of intensity modulated radiation therapy treatment plan. While QA procedures for motor-driven MLC have been published extensively, those for binary MLCs such as the one used for helical tomotherapy have not been presented in the literature, as this is still a fairly new technology. In this study, seven ...
متن کاملFirst experiences in using a dose control system on a TomoTherapy Hi·Art II
The purpose of this study was to investigate the impact of a dose control system (DCS) servo installed on two fully commissioned TomoTherapy Hi·Art II treatment units. This servo is designed to actively adjust machine parameters to control the output variation of a tomotherapy unit to within ± 0.5% of the nominal dose rate. Machine output, dose rate, and patient-specific quality assurance data ...
متن کاملPerformance Characteristics of an Independent Dose Verification Program for Helical Tomotherapy
Helical tomotherapy with its advanced method of intensity-modulated radiation therapy delivery has been used clinically for over 20 years. The standard delivery quality assurance procedure to measure the accuracy of delivered radiation dose from each treatment plan to a phantom is time-consuming. RadCalc®, a radiotherapy dose verification software, has released specifically for beta testing a m...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Physics in medicine and biology
دوره 49 13 شماره
صفحات -
تاریخ انتشار 2004