SU-E-T-642: Safety Procedures for Error Elimination in Cyberknife Stereotactic Radiosurgery (SRS) and Stereotactic Body Radiation Therapy (SBRT).

نویسندگان

  • A Hussain
  • A Alkafi
  • W Al-Najjar
  • B Moftah
چکیده

PURPOSE Cyberknife system is used for providing stereotactic radiosurgery (SRS) and stereotactic body radiation therapy (SBRT) hypofractionation scheme. The whole treatment delivery is based on live imaging of the patient. The minor error made at any stage may bring severe radiation injury to the patient or damage to the system itself. Several safety measures were taken to make the system safer. METHODS The radiation treatment provided thru a 6MV linac attached to Kuka robot (Cyberknife G4, Accuray Inc. Sunnyvale, CA, USA). Several possible errors were identified related to patient alignment, treatment planning, dose delivery and physics quality assurance. During dose delivery, manual and visual checks were introduced to confirm pre and intra-treatment imaging to reduce possible errors. One additional step was introduced to confirm that software tracking-tools had worked correctly with highest possible confidence level. Robotic head move in different orientations over and around the patient body, the rigidity of linac-head cover and other accessories was checked periodically. The vender was alerted when a tiny or bigger piece of equipment needed additional interlocked support. RESULTS As of our experience treating 525 patients on Cyberknife during the last four years, we saw on and off technical issues. During image acquisition, it was made essential to follow the site-specific imaging protocols. Adequate anatomy was contoured to document the respective doses. Followed by auto-segmentation, manual tweaking was performed on every structure. The calculation box was enclosing the whole image during the final calculation. Every plan was evaluated on slice-by slice basis. To review the whole process, a check list was maintained during the physics 2nd-check. CONCLUSION The implementation of manual and visual additional checks introduced along with automated checks for confirmation was found promising in terms of reduction in systematic errors and making the system safer than before.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Cost-Effectiveness Results

OBJECTIVE To describe and synthesize the current stereotactic radiosurgery (SRS) and stereotactic body radiation therapy (SBRT) cost-effectiveness research to date across several common SRS and SBRT applications. METHODS This review was limited to comparative economic evaluations of SRS, SBRT, and alternative treatments (e.g., other radiotherapy techniques or surgery). Based on PubMed searche...

متن کامل

CyberKnife Robotic Radiosurgery of Three Pancreatic Metastases of Renal Cell Cancer in One Treatment Session

Introduction: Pancreatic metastasis accounts for 2% to 11% of all renal cell cancer (RCC) metastases. In good clinical status, patients are selected for surgery, and in more advanced disease, for systemic chemotherapy. Stereotactic radiosurgery (SRS) and stereotactic body radiation therapy (SBRT) do not play a therapeutic role so far. Materials Methods: A 57-year-old patient was treated with Cy...

متن کامل

Stereotactic radiosurgery: a “targeted” therapy for cancer

The developments of medicine always follow innovations in science and technology. In the past decade, such innovations have made cancer-related targeted therapies possible. In general, the term "targeted therapy" has been used in reference to cellular and molecular level oriented therapies. However, improvements in the delivery and planning of traditional radiation therapy have also provided ...

متن کامل

Radiobiology of Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

References .......................................................................... 59 Abstract In recent years, increasing number of cancer patients are treated with stereotactic radiosurgery (SRS) or stereotactic body radiation therapy (SBRT), which deliver hypofractionated irradiation with high-dose per fraction . It is highly likely that the radiobiological principles such as 4 Rs (Reoxyg...

متن کامل

Effect of CyberKnife stereotactic body radiation therapy for hepatocellular carcinoma on hepatic toxicity

OBJECTIVE To evaluate the safety of CyberKnife stereotactic body radiation therapy (SBRT) for hepatocellular carcinoma (HCC) patients and identify the treatment-related risk factors of hepatic toxicity. MATERIALS AND METHODS One hundred and four HCC patients treated with CyberKnife SBRT were included in this study between August 2009 and December 2012. The average dose of prescribed radiation...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Medical physics

دوره 41 6  شماره 

صفحات  -

تاریخ انتشار 2014