Retrospective analysis of pulse oximeter alarm settings in an intensive care unit patient population
نویسندگان
چکیده
BACKGROUND The cacophony of alerts and alarms in a hospital produced by medical devices results in alarm fatigue. The pulse oximeter is one of the most common sources of alarms. One of the ways to reduce alarm rates is to adjust alarm settings at the bedside. This study is aimed to retrospectively examine individual pulse oximeter alarm settings on alarm rates and inter- and intra- patient variability. METHODS Nine hundred sixty-two previously collected intensive care unit (ICU) patient records were obtained from the Multiparameter Intelligent Monitoring in Intensive Care II Database (Beth Israel Deaconess Medical Center, Boston, MA). Inclusion criteria included patient records that contained SpO2 trend data sampled at 1 Hz for at least 1 h and a matching clinical record. SpO2 alarm rates were simulated by applying a range of thresholds (84, 86, 88, and 90 %) and delay times (10 to 60 s) to the SpO2 data. Patient records with at least 12 h of SpO2 data were examined for the variability in alarm rate over time. RESULTS Decreasing SpO2 thresholds and increasing delay times resulted in decreased alarm rates. A limited number of patient records accounted for most alarms, and this number increased as alarm settings loosened (the top 10 % of patient records were responsible for 57.4 % of all alarms at an SpO2 threshold of 90 % and 15 s delay and 81.6 % at an SpO2 threshold of 84 % and 45 s delay). Alarm rates were not consistent over time for individual patients with periods of high and low alarms for all alarm settings. CONCLUSION Pulse oximeter SpO2 alarm rates are variable between patients and over time, and the alarm rate and the extent of inter- and intra-patient variability can be affected by the alarm settings. Personalized alarm settings for a patient's current status may help to reduce alarm fatigue for nurses.
منابع مشابه
آگاهی کارکنان درمانی بخشهای ویژه نوزادان و کودکان درباره پالساکسیمتری
Background & Aim: Pulse Oximeter monitoring technology has become so common in intensive health care settings over the last decade that blood oxygen level is now considered as the fifth vital sign. However, it seems that medical and nursing staffs are not specially educated to operate with the devices. The aim of the study was to investigate the knowledge of medical assistants (residents), nurs...
متن کاملImproved Filtering of Pulse Oximeter Monitoring Alarms in the Neonatal ICU: Bedside Significance.
BACKGROUND The OxiMax N-600x containing SatSeconds alarm management software was designed to assist clinicians in discriminating nuisance alarms from those that are clinically relevant. Instead of sounding an alarm the moment the oxygen saturation reading violates the upper or lower limit settings, a magnitude and duration of tolerance can be set. The primary objective was to study the proporti...
متن کاملAssessing the Performance of Nurses in the Proper Adjustment of Monitoring Instruments in the Neonatal Intensive Care Unit
Background: Monitoring serves to maintain physiologic variables within normal limits and when a parameter crosses a set threshold, an alarm is triggered. Therefore, this study aimed to determine whether alarm limits are properly adjusted in the neonatal intensive care unit (NICU) by nursing staff or not.Methods: The data concerning alarm limits corresponding to the three shifts of nursing...
متن کاملThe effects of motion on the performance of pulse oximeters in volunteers (revised publication).
BACKGROUND Pulse oximetry is considered a standard of care in both the operating room and the postanesthetic care unit, and it is widely used in all critical care settings. Pulse oximeters may fail to provide valid SpO2 data in various situations that produce low signal-to-noise ratio. Motion artifact is a common cause of oximeter failure and loss of accuracy. This study compares the accuracy a...
متن کاملComparison of two new generation pulse oximeters with arterial oxygen saturation in critically ill children.
OBJECTIVES To compare the performance of two new generation pulse oximeters, one with enhanced signal extraction technology (SET) and other without enhanced SET in detecting hypoxemia and to correlate it with arterial blood gas analysis. METHODS Forty-eight patients, admitted to pediatric intensive care unit (PICU) of a teritiary care teaching hospital in India for critical care and support d...