Technical evaluation of different respiratory monitoring systems used for 4D CT acquisition under free breathing
نویسندگان
چکیده
Respiratory monitoring systems are required to supply CT scanners with information on the patient's breathing during the acquisition of a respiration-correlated computer tomography (RCCT), also referred to as 4D CT. The information a respiratory monitoring system has to provide to the CT scanner depends on the specific scanner. The purpose of this study is to compare two different respiratory monitoring systems (Anzai Respiratory Gating System; C-RAD Sentinel) with respect to their applicability in combination with an Aquilion Large Bore CT scanner from Toshiba. The scanner used in our clinic does not make use of the full time dependent breathing signal, but only single trigger pulses indicating the beginning of a new breathing cycle. Hence the attached respiratory monitoring system is expected to deliver accurate online trigger pulse for each breathing cycle. The accuracy of the trigger pulses sent to the CT scanner has to be ensured by the selected respiratory monitoring system. Since a trigger pulse (output signal) of a respiratory monitoring system is a function of the measured breathing signal (input signal), the typical clinical range of the input signal is estimated for both examined respiratory monitoring systems. Both systems are analyzed based on the following parameters: time resolution, signal amplitude, noise, signal-to-noise ratio (SNR), signal linearity, trigger compatibility, and clinical examples. The Anzai system shows a better SNR (≥ 28 dB) than the Sentinel system (≥ 14.6 dB). In terms of compatibility with the cycle-based image sorting algorithm of the Toshiba CT scanner, the Anzai system benefits from the possibility to generate cycle-based triggers, whereas the Sentinel system is only able to generate amplitude-based triggers. In clinical practice, the combination of a Toshiba CT scanner and the Anzai system will provide better results due to the compatibility of the image sorting and trigger release methods.
منابع مشابه
Reconstruction of 4D-CT from a Single Free-Breathing 3D-CT by Spatial-Temporal Image Registration
In the radiation therapy of lung cancer, a free-breathing 3D-CT image is usually acquired in the treatment day for image-guided patient setup, by registering with the free-breathing 3D-CT image acquired in the planning day. In this way, the optimal dose plan computed in the planning day can be transferred onto the treatment day for cancer radiotherapy. However, patient setup based on the simple...
متن کاملAn optical flow based method for improved reconstruction of 4D CT data sets acquired during free breathing.
Respiratory motion degrades anatomic position reproducibility and leads to issues affecting image acquisition, treatment planning, and radiation delivery. Four-dimensional (4D) computer tomography (CT) image acquisition can be used to measure the impact of organ motion and to explicitly account for respiratory motion during treatment planning and radiation delivery. Modern CT scanners can only ...
متن کاملEvaluation of 4D CT acquisition methods designed to reduce artifacts
Four-dimensional computed tomography (4D CT) is used to account for respiratory motion in radiation treatment planning, but artifacts resulting from the acquisition and postprocessing limit its accuracy. We investigated the efficacy of three experimental 4D CT acquisition methods to reduce artifacts in a prospective institutional review board approved study. Eighteen thoracic patients scheduled...
متن کاملImpact of incorporating visual biofeedback in 4D MRI
Precise radiation therapy (RT) for abdominal lesions is complicated by respiratory motion and suboptimal soft tissue contrast in 4D CT. 4D MRI offers improved con-trast although long scan times and irregular breathing patterns can be limiting. To address this, visual biofeedback (VBF) was introduced into 4D MRI. Ten volunteers were consented to an IRB-approved protocol. Prospective respiratory-...
متن کاملDevelopment of real‐time motion verification system using in‐room optical images for respiratory‐gated radiotherapy
Phase-based respiratory-gated radiotherapy relies on the reproducibility of patient breathing during the treatment. To monitor the positional reproducibility of patient breathing against a 4D CT simulation, we developed a real-time motion verification system (RMVS) using an optical tracking technology. The system in the treatment room was integrated with a real-time position management system. ...
متن کامل