The effect of pilot balloon design on estimation of safe tracheal tube cuff pressure.

نویسندگان

  • K M Janossy
  • J Pullen
  • D Young
  • G Bell
چکیده

We studied the effect of pilot balloon design on the ability of experienced anaesthetists to assess and inflate tracheal tube cuffs to safe pressures. A model trachea was designed, incorporating a degree of compliance and an air leak, to evaluate six different pilot balloons grafted onto identical tracheal tubes. Pilot balloons were inflated to one of four pressures and anaesthetists were asked to estimate whether the pressure was acceptable, too low or too high. Anaesthetists were then asked to inflate the cuff of each tube. Overall, 103 (42.9%) of anaesthetists' assessments of tracheal tube cuff pressures were correct (33% correct would be expected by chance, p = 0.002). Pressures generated by anaesthetists inflating tracheal tube cuffs were very variable. Median (IQR [range]) pressures for each pilot balloon ranged from 29 (17-43 [9-56]) cmH(2)O to 74 (49-114 [4-140]) cmH(2)O (p < 0.001). The design of the pilot balloon significantly affects anaesthetists' ability to inflate tracheal tube cuffs to safe pressures.

منابع مشابه

Comparison of tracheal tube cuff inflation pressure among anesthesia team with different gradual level

Background & Objectives: Tracheal intubation is frequently performed in operating room. Over inflation of the endotracheal tube cuff is a risk factor for tracheal ischemia and subsequent complications. On the other hand inadequate cuff pressure increases the risk of aspiration of gastric contents. In this study, the endotracheal tube cuff pressure filled by the anesthesia team has been compared...

متن کامل

Endotracheal Tube Cuff Pressure Assessment: Education May Improve but not Guarantee the Safety of Palpation Technique

BACKGROUND Endotracheal Tube Cuff Pressure (ETCP) should be kept in the range of 20 - 30 cm H2O. Earlier studies suggested that ETCP assessment by palpation of pilot balloon results in overinflation or underinflation and subsequent complications such as tracheal wall damage and aspiration. OBJECTIVES The current study aimed to evaluate the effect of an in vitro educational program on the abil...

متن کامل

Comparison of tracheal tube cuff pressure with two techniques: fixed volume versus pilot balloon palpation

Introduction: Filling tracheal tube cuff (TTC) after intubation is necessary to provide a safe airway in intubated patients. On the other hand, excessive increase or decrease in the pressure of TTC's balloon leads into the dangerous complications such as necrosis and/or aspiration. Accordingly, in the present study, we tried to evaluate the most two common fixed volume and pilot balloon palpita...

متن کامل

A comparison of endotracheal tube cuff pressures using estimation techniques and direct intracuff measurement.

Cuffed endotracheal tubes are one aspect of airway management designed to ensure safety, yet patients can be at risk for injury from underinflated and overinflated endotracheal cuffs. Tracheal pressures exceeding approximately 48 cm H2O impede capillary blood flow, potentially causing tracheal damage, and pressures below approximately 18 mm Hg may increase the risk of aspiration. There is no st...

متن کامل

بررسی تاثیر تغییر وضعیت بدن بر فشار کاف لوله تراشه در بیماران تحت تهویه مکانیکی

Background: Endotracheal tube cuff pressure must be maintained in safe range. Many factors could be affecting on endotracheal tube cuff pressure in patients on mechanical ventilation. Patients undergoing mechanical ventilation (MV) in critical care settings require changing position for different reasons. The aim of this study was to determine the effect of changes in body position and head of ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

متن کامل
عنوان ژورنال:
  • Anaesthesia

دوره 65 8  شماره 

صفحات  -

تاریخ انتشار 2010