Modeling photon output caused by backscattered radiation into the monitor chamber from collimator jaws using a Monte Carlo technique.

نویسندگان

  • H H Liu
  • T R Mackie
  • E C McCullough
چکیده

Dose per monitor unit in photon fields generated by clinical linear accelerators can be affected by the backscattered radiation into the monitor chamber from collimator jaws. Thus, it is necessary to account for the backscattered radiation in computing monitor unit setting for a treatment field. In this work, we investigated effects of the backscatter from collimator jaws based on Monte Carlo simulations of a clinical linear accelerator. The backscattered radiation scored within the monitor chamber was identified as originating either from the upper jaws (Y jaws), or from the lower jaws (X jaws). From the results of Monte Carlo simulations, ratios of the monitor-chamber-scored dose caused by the backscatter to the dose caused by the forward radiation, R(x,y), were modeled as functions of the individual X and Y jaw positions. The amount of the backscattered radiation for any field setting was then computed as a compound contribution from both the X and Y jaws. The dose ratios of R(x,y) were then used to calculate the change in photon output caused by the backscatter, Scb(x,y). Results of these calculations were compared with available measured data based on counting the electron pulses or charge from the electron target of an accelerator. Data from this study showed that the backscattered radiation contributes approximately 3% to the monitor-chamber-scored dose. A majority of the backscattered radiation comes from the upper jaws, which are located closer to the monitor chamber. The amount of the backscatter decreases approximately in a linear fashion with the jaw opening. This results in about a 2% increase of photon output from a 10 cm x 10 cm field to a 40 cm x 40 cm field. The off-axis location of the jaw opening does not have a significant effect on the magnitude of the backscatter. The backscatter effect is significant for monitor chambers using kapton windows, particularly for treatment fields using moving jaws. Applying the backscatter correction improves the accuracy of monitor-unit calculation using a model-based dose calculation algorithm such as the convolution method.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Interpretation of in-air output ratio and its impact on Dose Calculation

The concept of in-air output ratio (Sc) was introduced to characterize how the incident photon fluence per monitor unit (or unit time for a Co-60 unit) varies with collimator settings. 2 This quantity is also called inair output factor, collimator-scatter factor, and headscatter factor. 6 The latter two names were somewhat misleading since they emphasized a single component of the output ratio....

متن کامل

Comparison of measured and Monte Carlo calculated dose distributions from “circular collimators” for radiosurgical beams

Background: Stereotactic radiosurgery is an important clinical tool for the treatment of small lesions in the brain, including benign conditions, malignant and localized metastatic tumors. A dosimetry study was performed for Elekta ‘Synergy S ’ as a dedicated Stereotactic radiosurgery unit, capable of generating circular radiation fields with diameters of 1-5 cm at isocentre using the BEAM/EGS4...

متن کامل

A Monte Carlo Simulation of Photon Beam Generated by a Linear Accelerator

ntroduction:  Monte  Carlo  simulation  is  the  most  accurate  method  of  simulating  radiation  transport  and  predicting doses at different points of interest in radiotherapy. A great advantage of the Monte Carlo method  compared  to  the  deterministic  methods  is  the  ability  to  deal  accurately  with  any  complex  geometry.  Its  disadvantage is the extremely long computing time r...

متن کامل

Monte Carlo Modeling of Dual and Triple Photon Energy Absorptiometry Technique

Introduction: Osteoporosis is a bone disease in which there is a reduction in the amount of bone mineral content leading to an increase in the risk of bone fractures. The affected individuals not only have to go through lots of pain and suffering but this disease also results in high economic costs to the society due to a large number of fractures.  A timely and accurate diagnosis of this disea...

متن کامل

Assessment of Effect Technical Directional Bremsstrahlung Splitting (DBS) on Spectra and Parameters of Simulation with Monte carlo Method BEAMnrc Code (Study Monte Carlo)

Introduction: Previous studies have shown that a Monte Carlo method for the transportations photon beam in medical linear accelerator is a good way. Strip of simulation can be used to measure the dose distribution in phantoms and patients' body. EGSnrc Code is the only code written for use in the field of radiation therapy that has many subset codes that BEAMnrc code is an impo...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Medical physics

دوره 27 4  شماره 

صفحات  -

تاریخ انتشار 2000