Results of selfmonitoring on glucometer systems Advance and Optium in daily routine.

نویسندگان

  • Rudolf Chlup
  • Martin Payne
  • Jana Zapletalová
  • Stanislav Komenda
  • Blanka Doubravová
  • Martina Reznícková
  • Ludmila Chlupová
  • Pavel Seckar
چکیده

The aim of this prospective clinical study was to compare the results of B-glucose estimations performed simultaneously on glucometer Advance (with Micro-draw strips) and Optium (G3 strips) by lay healthy volunteers under non-standardized conditions of everyday life, to assess the difficulties dealing with lay-handling of these systems and to demonstrate the possibilities of the software Glucobalance (Hypoguard) and PC-Link (Medisense/Abbott) for the analysis of selfmonitoring. In the course of 5 days, a total of 721 pairs of measurements were carried out on 10 pairs of glucometer Advance and Optium by 10 healthy volunteers aged 16-40 years. The data transfer of all values into computer from glucometer Advance using the Glucobalance software and from glucometer Optium using the PC-Link was carried out to determine the results. The correlation of B-glucose measured on the glucometer Advance and Optium was strong (r = 0.73). Glucometer Advance brings values about 0.21 +/- 0.06 mmol/l lower than glucometer Optium. The average difference found within each pairs of glucometers Advance - Optium varied. Nevertheless, these differences are acceptable for routine selfmonitoring. The handling of glucometer Advance is not difficult for lay persons. The Glucobalance software simplifies the result evaluation by each tested person. Even though there are some advantages in comparison with the PC-Link, it should be further developed.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Effect of fluoride-containing tubes on accuracy of glucometers.

G lucometers are used to quickly obtain glucose concentrations in the hospital and at home. For quality assurance, the American Diabetes Association recommends periodically comparing the results from glucometers with those from referenced laboratories (1). Usually, measurement using glucometers is done one-by-one by technicians, which is costly regarding manpower. In contrast, measurement using...

متن کامل

مقایسه قدرت دستگاه گلوکومتر Accu-chek با Elite در تشخیص دیابت با استفاده از خونریزی متعاقب Probing

Abstract Background: Recent studies have assessed the validity of glucometer devices in estimating the glucose level of gingival blood during routine periodontal examination (Bleeding On Probing). The purpose of this study was to evaluate whether blood oozing from gingival tissues during routine periodontal examination can be used for determining g...

متن کامل

مقایسه میزان همخوانی یافته‌های سنجش قندخون با استفاده از گلوکومتر ON CALL و گلوکوکارد با روش آزمایشگاهی استاندارد

Background: Diabetes Mellitus (DM) is one the most common chronic disease, with many complications including renal failure, blindness and non traumatic amputation. Prevention of complications is possible through monitoring and control of blood glucose levels. Considering how easy home blood glucose monitoring is, we decided to evaluate the performance of two available glucometers for detection...

متن کامل

Optimization of two-stage production/inventory systems under order base stock policy with advance demand information

It is important to share demand information among the members in supply chains. In recent years, production and inventory systems with advance demand information (ADI) have been discussed, where advance demand information means the information of demand which the decision maker obtains before the corresponding actual demand arrives. Appropriate production and inventory control using demand info...

متن کامل

A Comparative Study of Blood Glucose Measurements Using Glucometer Readings and the Standard Method in the Diagnosis of Neonatal Hypoglycemia

Background: Hypoglycemia is one of the most common neonatal disorders, associated with severe complications. There has been a great deal of controversy regarding the definition and screening of hypoglycemia. Therefore, in this study, we aimed to determine a cut-off value for blood glucose level in glucometer readings. Methods: This cross-sectional study was conducted on 238 newborns at risk of ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Biomedical papers of the Medical Faculty of the University Palacky, Olomouc, Czechoslovakia

دوره 149 1  شماره 

صفحات  -

تاریخ انتشار 2005