Breast Diffusion-weighted Imaging at 3 and 7 Tesla: Comparison Study

نویسندگان

  • Lenka Minarikova
  • Wolfgang Bogner
  • Olgica Zaric
  • Katja Pinker-Domenig
  • Thomas Helbich
  • Siegfried Trattnig
  • Stephan Gruber
چکیده

PURPOSE: Diffusion weighted imaging is a potential method for the detection and characterization of breast cancer [1]. By measuring apparent diffusion coefficient (ADC) values, it is possible to probe the local cellular microstructure in breast tissue. This may be used to distinguish between malignant and benign lesions and to recognize early response in cancer therapy monitoring. High field MR systems offer a potential gain in sensitivity that can be translated to higher spatial resolution. This could improve lesion detection and differentiation. The purpose of this study was to compare bilateral breast-DWI with sub-mm-resolution (0.9mm) at 7T with 3T (1.4mm) in the same patients. METHODS: All experiments were performed at 3T and 7T MRI systems (3T Trio, 7T Magnetom, Siemens Healthcare, Erlangen, Germany) using dedicated bilateral breast coils with four 1H-channels (3T: In vivo, Orlando, FL, USA; 7T: 1H/31P-coil, H. Stark, Erlangen, Germany). This study was approved by the local ethics committee. 15 patients (55+/-17yo) were measured at both field strengths within one week. In pre-menopausal women, MRI was performed between day 7 and 14 of the menstrual cycle. Bilateral 3 direction trace DWI was performed in 2:56min (3T) and 3:48min (7T) using combined readoutsegmented echo planar imaging (rs-EPI) with fat-suppression and with b-values of 0 and 850 sec/mm. The in-plane resolution was 1.4x1.4mm at 3T and 0.9x0.9mm2 at 7T with the same slice thickness (5mm). To allow a better resolution at 7T without causing additional blurring or distortions, we used 9 readout segments at 7T compared to 5 at 3T. To compensate for the expected increase in offresonance effects at 7T, we additionally used parallel imaging at 7T (GRAPPA, factor 2). In post-processing, two-dimensional regions of interest (ROIs) were drawn manually in the lesion, and in homogeneous breast parenchyma without enhancement in the contralateral breast in three different parts of the breast: paramammilar, central and pre-pectoral, using OsiriX® software. SNR was defined as the ratio between mean signal amplitude inside the ROI (SROI) and standard deviation of the background noise (σBG) (SNR=SROI/σBG). The statistical analysis was performed in IBM SPSS Statistics (Armonk, NY). For the analysis, Student’s t-test was used for healthy tissue data, because of possible difference in partial volume effects in ROIs in lower density breast cases. A paired test was used for benign and malignant lesions data. RESULTS: 15 lesions were identified – 7 benign and 8 malignant. Benign lesions were all fibroadenomas. In malignant cases, there were 7 invasive ductal carcinomas and one invasive lobular carcinoma. There was no significant difference between SNR in different regions of breast (3T: p=0.306, 7T: p=0.086) and neither between the 3T and 7T (p = 0.841). The SNR comparison can be seen as boxplots in Figure 3. ADC values in normal tissue were not significantly different between the various regions within the breast (3T: p=0.924; 7T: p=0.622). The boxplot comparison of ADC values at 3T and 7T in different lesion types and normal tissue can be seen in Figure 1. The average ADC values can be found in the table in Figure 2. They were found to be different for different field strengths, in two cases: in healthy tissue and in malignant lesion ROIs (paired values).

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

[Examination of upper abdominal region in high spatial resolution diffusion-weighted imaging using 3-Tesla MRI].

The advantage of the higher signal-to-noise ratio (SNR) of 3-Tesla magnetic resonance imaging (3-Tesla) has the possibility of contributing to the improvement of high spatial resolution without causing image deterioration. In this study, we compared SNR and the apparent diffusion coefficient (ADC) value with 3-Tesla as the condition in the diffusion-weighted image (DWI) parameter of the 1.5-Tes...

متن کامل

The role of apparent diffusion coefficient values in the differential diagnosis of breast lesions in diffusion-weighted MRI.

PURPOSE We aimed to determine the apparent diffusion coefficients (ADCs) of focal breast lesions on diffusion-weighted magnetic resonance imaging (DW-MRI) and to evaluate whether ADC measurement can be used to characterize lesions as benign or malignant. MATERIALS AND METHODS Fifty-one patients between the ages of 18-79 years (mean age, 48.5 years) with 51 histopathologically verified breast ...

متن کامل

Differentiation between Benign and Malignant Breast Lesions Using ADC on Diffusion-Weighted Imaging at 3.0 T

Aim: To investigate the diagnostic efficiency of apparent diffusion coefficient value (ADC) in differentiating benign from malignant breast lesions at 3.0 T diffusion-weighted imaging with different pair of b value. Methods: Total 110 patients with 107 lesions (44 benign and 63 malignant) were selected for our study with five different b-values 0, 400, 800, 1200 and 1600 s/mm2. ADC values were ...

متن کامل

1.5 versus 3 versus 7 Tesla in abdominal MRI: A comparative study

OBJECTIVES The aim of this study was to investigate and compare the feasibility as well as potential impact of altered magnetic field properties on image quality and potential artifacts of 1.5 Tesla, 3 Tesla and 7 Tesla non-enhanced abdominal MRI. MATERIALS AND METHODS Magnetic Resonance (MR) imaging of the upper abdomen was performed in 10 healthy volunteers on a 1.5 Tesla, a 3 Tesla and a 7...

متن کامل

Diffusion Weighted Imaging at 7T with STEAM-EPI and GRAPPA

Introduction: Diffusion Imaging at 7 Tesla has been quite a challenge. Standard diffusion sequences require long time to play diffusion gradients, especially for high b-values. This compromises higher intrinsic SNR [1]at 7T with shorter T2 relaxation. At 7T twice refocused spin echo [2] used for standard diffusion weighted imaging also suffers from high SAR. Magnetic field inhomogeneities that ...

متن کامل

Diagnostic Value of Diffusion-weighted Imaging and Apparent Diffusion Coefficient Values in the Differentiation of Breast Lesions, Histpathologic Subgroups and Correlatıon with Prognostıc Factors using 3.0 Tesla MR.

OBJECTIVE The aim of this study was to evaluate the effect of the apparent diffusion coefficient (ADC) and diffusion-weighted imaging in differentiating benign from malignant breast lesions, histopathologic subtypes of breast tumors, and to find a correlation with prognostic factors using 3T MR. MATERIALS AND METHODS A total of 165 patients aged between 16 and 78 years with 181 histopathologi...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2013