Nonenzymatic stopped-flow fluorimetric method for direct determination of uric acid in serum and urine.
نویسندگان
چکیده
A simple, direct, sensitive, and selective stopped-flow method for the fluorimetric determination of uric acid in serum and urine samples is described. The variation of fluorescence intensity during the reaction between uric acid and 1,1,3-tricyano-2-amino-1-propene (triap) in the presence of hydrogen peroxide is monitored. For these kinetic measurements peroxide is monitored. For these kinetic measurements we used a versatile stopped-flow module that can be fitted to any fluorimeter. The linear range of the proposed method is 0.08-3.00 mg of uric acid per liter, and the detection limit is 0.03 mg/L. Within- and between-assay CVs and selectivity data are reported. Results for serum and urine samples correlated well with those obtained by the uricase method The proposed method is inexpensive and requires no sophisticated detection equipment.
منابع مشابه
A Sensitive Simultaneous Determination of Uric Acid, Norepinephrine and Indomethacin using a Cadmium Sulfide Nanoparticles/Multi-Walled Carbon Nanotubes Modified Gold Electrode
In this study a novel method was developed to fabricate cadmium sulfide nanoparticles/multiwalled carbon nanotubes composite modified gold electrode (CdSNPs/MWCNTS/AuE) to measure trace amounts of norepinephrine(NE), indomethacin (IND), and uric acid (UA) simultaneously. Electrochemical investigations were carried out using cyclic voltammetry (CV), differential pulse voltammetry (DPV) and chron...
متن کاملSimultaneous determination of dopamine and uric acid using a glassy carbon paste electrode modified with copper- para red complex
A simple approach based on cyclic voltammetry (CV) was developed for the simultaneous determination of dopamine (DA) and uric acid (UA) in the presence of ascorbic acid (AA) using a modified glassy carbon paste electrode (GCPE). In the present study, analytical parameters were optimized and electrochemical performance of modified electrode was investigated. The calibration curves were obtained ...
متن کاملAn efficient platform based on cupper complex-multiwalled carbon nanotube nanocomposite modified electrode for the determination of uric acid
A new voltammetric sensor for determination of uric acid (UA) by Cuppercomplex- multiwalled carbon nanotube (Cu-complex-CNT) nanocomposite modifiedcarbon paste electrode (CPE) is reported. The electrocatalytic behavior of theCu-complex-CNT nanocomposite modified CPE was studied in pH 2.0 phosphatebuffer solution by chronoamperometry (CA) and cyclic voltammetry (CV) in th...
متن کاملFluorimetric determination of total ascorbic acid by a stopped-flow mixing technique.
A simple, rapid and automatic fluorimetric method for the determination of total ascorbic acid is described. The method makes use of the stopped-flow mixing technique in order to achieve the rapid oxidation of ascorbic acid by dissolved oxygen to dehydroascorbic acid, which then reacts with o-phenylenediamine to form a fluorescent quinoxaline. The initial rate and fluorescence signal of this sy...
متن کاملSimultaneous Voltammetric Measurement of Ascorbic Acid, Epinephrine, Uric Acid and Tyrosine at a Glassy Carbon Electrode Modified with Nanozeolite-Multiwall Carbon Nanotube
In this study, incorporation of iron ion-doped natrolite nanozeolite, multi-wall carbon nanotubes into chitosan-coated glassy carbon electrode for the simultaneous determination of ascorbic acid, epinephrine, uric acid and tyrosine is studied. The results show that the combination of multi-wall carbon nanotubes and iron ion-doped natrolite zeolite causes a dramatic enhancement in the sensitivit...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinical chemistry
دوره 35 2 شماره
صفحات -
تاریخ انتشار 1989