Evaluation of the Olympus Demand random access chemistry analyser
نویسندگان
چکیده
We evaluated the Olympus Demand as a cost-saving measure, as a back up to a Technicon SMAC I, and to be the primary instrument for enzyme analysis. In general, it wasfound to be precise, reliable, easy to operate, but with only average turn-around time capabilities. Technically there were several limitations: (1) significant bilirubin interference in the analysis of serum bicarbonate and creatinine; (2) instability ofthe ion selective electrodes; (3) lengthy routine maintenance requiredfor the ion selective electrode module; (4) serum phosphorus and uric acid methods required blanking due to interferencefrom bilirubin, hemolysis, or lipemia; and (5) serum triglyceride analysis included measurement offree glycerol.
منابع مشابه
A systematic evaluation of the Olympus AU5061 as an effective replacement for the SMAC II analyser
متن کامل
Laboratory evaluation of the COBAS MIRA S random access analyser
The random access analyser COBAS MIRA S (Roche Diagnostics) was evaluated for two months. The instrument is a computer-controlled discrete analyser which can be run in a combination profile and single test mode. This instrument has special features, including an automatic cuvette segment changer, a reagent rack cooling system, an external keyboard and monitor, as well as a bar-code facility for...
متن کاملEvaluation of the Olympus AU-510 analyser
The selective multitest Olympus AU-510 analyser was evaluated according to the recommendations of the Comision de Instrumentacion de la Sociedad Española de Quimica Clinica and the European Committee for Clinical Laboratory Standards. The evaluation was carried out in two stages: an examination of the analytical units and then an evaluation in routine work conditions. The operational characteri...
متن کاملAnalytical performance of the selective, automatic multianalyser Olympus AU 5031
The analytical performance of the selective, automatic multianalyser Olympus AU 5031 was evaluated over four months and assessed for practicability for another eight months. The evaluation followed the ECCLS guidelines. Twenty routine parameters were measured. In addition, sodium and potassium were determined on an attached flame photometric unit. Both the agreement between the eights photomete...
متن کاملLetter to the Editor: The effects of hyperlipidaemia, hyperbilirubinaemia and haemolysis on tests performed by the Olympus AU 5000 multiple analyser
Increasing amounts ofIntralipid (Kabivitrum), bilirubin (NBS standard 916) or haemolysate were added to a serum pool of 40 patients, as previously described [1]. Added Intralipid yielded samples with triglyceride levels of 0"8-6-3 g/l, added bilirubin yielded samples having total bilirubin levels of 4-325 mg/1 and added haemolysate yielded samples with haemoglobin concentrations of 0-1"58 g/1. ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- The Journal of Automatic Chemistry
دوره 9 شماره
صفحات -
تاریخ انتشار 1987