Accuracy of five electronic foramen locators with different operating systems: an ex vivo study

نویسندگان

  • Bruno Carvalho de VASCONCELOS
  • Michelli de Medeiros BUENO
  • Suyane Maria LUNA-CRUZ
  • Marco Antonio Hungaro DUARTE
  • Carlos Augusto de Oliveira FERNANDES
چکیده

OBJECTIVE The aim of this study was to evaluate, ex vivo, the precision of five electronic root canal length measurement devices (ERCLMDs) with different operating systems: the Root ZX, Mini Apex Locator, Propex II, iPex, and RomiApex A-15, and the possible influence of the positioning of the instrument tips short of the apical foramen. MATERIAL AND METHODS Forty-two mandibular bicuspids had their real canal lengths (RL) previously determined. Electronic measurements were performed 1.0 mm short of the apical foramen (-1.0), followed by measurements at the apical foramen (0.0). The data resulting from the comparison of the ERCLMD measurements and the RL were evaluated by the Wilcoxon and Friedman tests at a significance level of 5%. RESULTS Considering the measurements performed at 0.0 and -1.0, the precision rates for the ERCLMDs were: 73.5% and 47.1% (Root ZX), 73.5% and 55.9% (Mini Apex Locator), 67.6% and 41.1% (Propex II), 61.7% and 44.1% (iPex), and 79.4% and 44.1% (RomiApex A-15), respectively, considering ±0.5 mm of tolerance. Regarding the mean discrepancies, no differences were observed at 0.0; however, in the measurements at -1.0, the iPex, a multi-frequency ERCLMD, had significantly more discrepant readings short of the apical foramen than the other devices, except for the Propex II, which had intermediate results. When the ERCLMDs measurements at -1.0 were compared with those at 0.0, the Propex II, iPex and RomiApex A-15 presented significantly higher discrepancies in their readings. CONCLUSIONS Under the conditions of the present study, all the ERCLMDs provided acceptable measurements at the 0.0 position. However, at the -1.0 position, the ERCLMDs had a lower precision, with statistically significant differences for the Propex II, iPex, and RomiApex A-15.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The effect of different needle-insertion depths on the accuracy of the Root ZX II and Root ZX mini apex locators in the presence of various irrigants

BACKGROUND AND AIM: The aim of the present study was to evaluate the effect of different needle-insertion depths on the accuracy of the Root ZX II and Root ZX mini apex locators in the presence of various irrigants. METHODS: Ninety extracted single-canal human teeth were used in nine experimental groups. The coronal one-third of the canals was preflared, and the length to the major foramen was ...

متن کامل

In vitro evaluation of the accuracy of two new electronic apex locators in root canal length determination

In vitro evaluation of the accuracy of two new electronic apex locators in root canal length determination Dr. Hasheminia, S.M.* - Dr. Jafari, N.** *Associate Professor of Endodontics Dept., Faculty of Dentistry, Isfahan University of Medical Sciences. ** Dentist. Abstract Background & Aim: Radiography is still the common method of measuring working length in root canal therapy. While, gag refl...

متن کامل

Ex vivo evaluation of the accuracy of two electronic apex locators during root canal length determination in primary teeth.

AIM To evaluate ex vivo the accuracy of two electronic apex locators during root canal length determination in primary incisor and molar teeth with different stages of physiological root resorption. METHODOLOGY One calibrated examiner determined the root canal length in 17 primary incisors and 16 primary molars (total of 57 root canals) with different stages of root resorption based on the ac...

متن کامل

A comparison between two electronic apex locators: an ex vivo investigation.

AIM To compare ex vivo the performance of the Apex Finder and the Root ZX apex locators, with and without irrigant, in canals having different diameters. METHODOLOGY Sixty canals in 60 teeth were prepared using stainless steel hand files and 0.04 taper NiTi rotary instruments. During preparation the narrowest diameter of the canal was transported to the apical root surface. The canals were ir...

متن کامل

Accuracy of three electronic apex locators in anterior and posterior teeth: an ex vivo study.

INTRODUCTION The aim of this study was to determine in anterior teeth, bicuspids, and molars (1) the accuracy of 3 different electronic apex locators (EALs) in detecting the apical foramen and (2) the accuracy of digital radiography in determining the working length (WL), compared with visible control under a microscope. METHODS By using radiovideography (RVG), we measured the lengths of 120 ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 21  شماره 

صفحات  -

تاریخ انتشار 2013