OBJECTIVES The present study evaluates complete supra-annular bioprostheses in patients with an aortic annulus of 18 to 23 mm in diameter. BACKGROUND Aortic valve replacement in patients with small aortic annulus using stented bioprostheses is often associated with unsatisfactory hemodynamic results and high incidence of patient-prosthesis mismatch. METHODS Between February 2000 and January 2004, 156 patients with aortic valve disease and an aortic annulus of 18 to 23 mm in diameter received the stented bovine Soprano (Sorin Biomedica Cardio, Saluggia, Italy) (n = 18), Perimount (Edwards Lifesciences, Irvine, California) (n = 52), Perimount Magna (Edwards Lifesciences) (n = 42), or the stented porcine Mosaic (Medtronic Inc., Minneapolis, Minnesota) (n = 44) bioprostheses. Intraoperatively, the surgeon measured the aortic annulus diameter by inserting a hegar dilator. Thus, postoperative hemodynamic results could be referred to the patient's aortic annulus diameter instead of referring the results to the labeled valve size. This allows for objective comparisons between different valve types. RESULTS There was no significant difference in hemodynamic results between the different valve types in patients with an aortic annulus 18 to 20 mm. In patients with an annulus 21 to 23 mm, the Magna was significantly superior to the other investigated devices in mean pressure gradient, effective orifice area, and incidence of patient-prosthesis mismatch. There was no significant difference between the complete supra-annular bioprostheses Mosaic and Soprano and the intra-supra-annular Perimount valve. CONCLUSIONS In patients with an aortic annulus of 18 to 20 mm in diameter, hemodynamic performance is independent of the implanted stented valve type and the annular position. Root enlargement or stentless valves may be beneficial alternatives. Patients with annulus diameter 21 to 23 mm benefit from the Magna in complete supra-annular position leading to superior hemodynamic results.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The complete supraannular concept: in vivo hemodynamics of bovine and porcine aortic bioprostheses.

BACKGROUND Complete supraannular placement of an aortic bioprosthesis is one approach to optimize the hemodynamic result of an aortic valve replacement. It is achieved with the combination of a special valve design and the supraannular sewing technique with noneverted mattress sutures. We evaluated 5 bioprostheses designed for complete supraannular placement to assess potential hemodynamic diff...

متن کامل

Totally Supra-Annular Mechanical Valve: to provide good hemodynamic performance and good clinical outcomes

When replacing an aortic valve with a mechanical prosthesis it is essential to make the right decision, mine was on the Bicarbon Overline valve (SORIN Group). Bicarbon Overline is a bileaflet mechanical valve designed for a Totally Supra-Annular implant. The innovative design of the valve is highlighted by curved leaflets with an 80° opening angle and an aerofoil titanium housing profile leadin...

متن کامل

[A comparative study of the follow-up and hemodynamics in vivo of 21 mm Carpentier-Edwards supra-annular and perimount bioprostheses].

INTRODUCTION AND OBJECTIVES Analysis and comparison of the clinical performance and hemodynamics in vivo of 21 mm Carpentier-Edwards supra-annular (CESA) and Perimount (CEPM) aortic bioprostheses. METHODS A follow-up study was made of 40 patients implanted a 21 mm CESA (n = 21) or CEPM (n = 19) prosthesis between October 1992 and September 1997. All eligible survivors (14 CESA, 12 CEPM) were ...

متن کامل

Comparison of eight prosthetic aortic valves in a cadaver model.

OBJECTIVES Proper valve selection is critical to ensure appropriate valve replacement for patients, because implantation of a small valve might place the patient at risk for persistent gradients. Labeled valve size is not the same as millimeter measure of prosthetic valve diameters or the annulus into which it will fit. Studies that use the labeled valve size in lieu of actual measured diameter...

متن کامل

Cardiac bioprostheses in the 1990s.

T he first successful heart valve replacements were carried out in 1960 with the use of mechanical prostheses.1-3 Because of the risk of throm-boembolism that is inherent to all mechanical prosthe-ses and the hazards of continuous anticoagulation, fewer thrombogenic tissue valves were developed.4,5 In the early 1960s, homograft replacement of the aortic valve was reported,6'7 and shortly therea...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of the American College of Cardiology

دوره 45 12  شماره 

صفحات  -

تاریخ انتشار 2005