Assessment of the Nova StatSensor whole blood point-of-care creatinine analyzer for the measurement of kidney function in screening for chronic kidney disease.
نویسندگان
چکیده
BACKGROUND Point-of-care testing for creatinine using a fingerprick sample and resultant estimated glomerular filtration rate has potential for screening for chronic kidney disease in community settings. This study assessed the applicability of the Nova StatSensor creatinine analyzer for this purpose. METHODS Fingerprick samples from 100 patients (63 renal, 37 healthy volunteers; range 46-962 micromol/L) were assayed using two StatSensor analyzers. Lithium heparin venous plasma samples collected simultaneously were assayed in duplicate using the isotope dilution mass spectrometry-aligned Roche Creatinine Plus enzymatic assay on a Hitachi Modular P unit. Method comparison statistics and the ability of the StatSensor to correctly categorise estimated glomerular filtration rate above or below 60 mL/min were calculated pre- and post-alignment with the laboratory method. RESULTS StatSensor 1 creatinine results (y) were much lower than the laboratory (y=0.75x+10.2, average bias -47.3, 95% limits of agreement -208 to +113 micromol/L). For estimated glomerular filtration rates above or below 60 mL/min, 100% and 87% of results respectively agreed with the laboratory estimated glomerular filtration rate (79% and 96% post-alignment). StatSensor 2 statistics were similar. The 95% limits of agreement between StatSensor creatinine results were -35 to +34 micromol/L. CONCLUSIONS Isotope dilution mass spectrometry alignment of the StatSensor will identify most patients with estimated glomerular filtration rate <60 mL/min, but there will be many falsely low estimated glomerular filtration rate results that require laboratory validation. Creatinine results need improvement.
منابع مشابه
Evaluation of the Nova StatSensor® XpressTM Creatinine Point-Of-Care Handheld Analyzer
Creatinine is a parameter that is required to monitor renal function and is important to follow in patients under treatment with potentially toxic renal drugs, such as the anti-HIV drug Tenofovir. A point of care instrument to measure creatinine would be useful for patients monitoring in resource-limited settings, where more instruments that are sophisticated are not available. The StatSensor X...
متن کاملEvaluation of Glomerular Filtration Rate Estimating Formulas in Diabetic Patients with Chronic Kidney Disease
Objective: Measurement of glomerular filtration rate (GFR) is the best determinant in assessment of kidney function for diagnostic and therapeutic purposes. As extremely accurate methods of GFR measurement (i.e. Inulin clearance) are expensive and Time-consuming and due to limitations of the 24 hour urine collection method, some formula have been developed for the GFR measurement. Here we have ...
متن کاملLong term effects of kidney donation on blood pressure and kidney function of donors
Introduction: Kidney transplantation from a live donor has many advantages, but the donor should undergo a major surgery and accept potential morbidity and mortality. In this study, we evaluated the long term effects of nephrectomy on renal function and blood pressure of kidney donors. Methods: In this cross-sectional study, we assessed 195 kidney donors. The controls were 100 healthy individu...
متن کاملEvaluation of Salivary Glucose and Creatinine for Screening of Diabetes and Kidney Disease
Objective: Diabetes mellitus is a multifactorial metabolic disorder that affects multiple organs of the body, one of the most important of which is chronic kidney disease. Early diagnosis of this disease is suggested as an important strategy to reduce the complications of the disease worldwide. Since the measurement of salivary creatinine and glucose is an easy and noninvasive method, the aim o...
متن کاملCorrelation between the Level of Creatinine, Serum Cardiac Troponines and Left Ventricle Function Tests in Patients with Chronic Renal Diseases without Acute Coronary Syndrome
Background and Objectives: The aim of this study was to survey the relationship between the level of creatinine,cardiac troponins serum and the function of the left ventricle in patients with kidney insufficiency,without acute coronary syndrome. Materialsand Methods:The sample was150 patients with nondialysis chronic kidney disease without acute coronary syndrome, hospitalized at I...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinical chemistry and laboratory medicine
دوره 48 8 شماره
صفحات -
تاریخ انتشار 2010