Low temperature thermometry in high magnetic fields. VII. CernoxTM sensors to 32 T
نویسندگان
چکیده
This article describes the behavior of CernoxTM zirconium oxynitride film temperature sensors from Lake Shore Cryotronics, Inc. in magnetic fields up to 32 T at temperatures between 2 and 286 K. Results from a number of sensors with different dimensionless temperature sensitivities and from different production batches are analyzed and compared with previous results on carbon-glass sensors that extend to 19 T. In that field range, the CernoxTM sensors appear to be a real alternative to carbon glass. Furthermore, the magnitude of their uncorrected error, DT/T , is smaller than other sensors at most temperatures in fields less than 20 T, and their temperature correctabilities appear to apply to off-the-shelf sensors with dimensionless temperature sensitivities in the range of 20.74 to 21.9. The sensors show a negligible ~,0.05%! orientation dependence of their DR/R at 78 K; at 4.2 K, that dependence can be as high as ;0.7% at 20 T. © 1999 American Institute of Physics. @S0034-6748~99!04701-2#
منابع مشابه
Integrated electronic transport and thermometry at milliKelvin temperatures and in strong magnetic fields.
We fabricated a He-3 immersion cell for transport measurements of semiconductor nanostructures at ultra low temperatures and in strong magnetic fields. We have a new scheme of field-independent thermometry based on quartz tuning fork Helium-3 viscometry which monitors the local temperature of the sample's environment in real time. The operation and measurement circuitry of the quartz viscometer...
متن کاملNoise Equivalent Power Optimization of Graphene- Superconductor Optical Sensors in the Current Bias Mode
In this paper, the noise equivalent power (NEP) of an optical sensor based ongraphene-superconductor junctions in the constant current mode of operation has beencalculated. Furthermore, the necessary investigations to optimize the device noise withrespect to various parameters such as the operating temperature, magnetic field, deviceresistance, voltage and current bias have been presented. By s...
متن کاملAccelerated MRI thermometry by direct estimation of temperature from undersampled k-space data.
PURPOSE Acceleration of magnetic resonance (MR) thermometry is desirable for several applications of MR-guided focused ultrasound, such as those requiring greater volume coverage, higher spatial resolution, or higher frame rates. METHODS We propose and validate a constrained reconstruction method that estimates focal temperature changes directly from k-space without spatial or temporal regula...
متن کاملTowards Motion-Robust Magnetic Resonance Thermometry
Magnetic Resonance Imaging allows for minimally invasive targeting and thermal ablation of tumors while monitoring the temperature distribution. Conventional MR thermometry procedures are hampered by either low accuracy or high sensitivity to motion artifacts due to the use of a reference temperature image. A new, dual-echo technique has been developed to obtain the temperature distribution wit...
متن کاملPCA-Based Magnetic Field Modeling : Application for On-Line MR Temperature Monitoring
Magnetic Resonance (MR) temperature mapping can be used to monitor temperature changes during minimally invasive thermal therapies. However, MR-thermometry contains artefacts caused by phase errors induced by organ motion in inhomogeneous magnetic fields. This paper proposes a novel correction strategy based on a Principal Component Analysis (PCA) to estimate magnetic field perturbation assumin...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره شماره
صفحات -
تاریخ انتشار 1998