Implant-abutment screw joint preload of 7 hex-top abutment systems.

نویسندگان

  • K B Tan
  • J I Nicholls
چکیده

This study measured the screw joint preload of the implant-abutment interface for 7 common hex-top abutment systems. Increasingly, prosthetic designs are utilizing a direct prosthetic connection to the implant, placing the implant-abutment screw joint under direct functional loads and moments. Sufficiently high screw joint preloads are required to maintain screw joint integrity and confer clinical longevity to implant prosthetic components to prevent such complications as abutment screw loosening and screw fracture. Strain-gauged abutment load cells were calibrated to measure screw joint pre-load at the implant-abutment interface. Torque delivery by electronic torque controller was varied at low- and high-speed settings. At manufacturer's recommended torques, the overall mean preload measured was 181.6 +/- 60.0 N for the Nobel Biocare Standard abutment, 291.3 +/- 41.2 N for the Nobel Biocare EsthetiCone abutment, 456.5 +/- 44.0 N for the Nobel Biocare MirusCone abutment, 369.7 +/- 32.9 N for the 3i Titanium Abutment post, 643.4 +/- 143.1 N for the Nobel Biocare CeraOne abutment, 536.3 +/- 68.6 N for the Nobel Biocare "Gold Cylinder to Fixture" abutment, and 556.9 +/- 145.6 N for the Nobel Biocare TiAdapt abutment. Analysis of variance revealed significant differences between abutment systems (P < .001) and between torque driver speed settings (P < .001). Implant-abutment screw joint preload of external-hex implants is dependent on abutment design, screw diameter, material, tightening torque, and torque controller speed.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Influence of Occlusal Forces on Stress Distribution on Preloaded Dental Implant Abutment Screws: A Finite Element Analysis Study

Purpose: The aim of the study is to determine stress distribution on preloaded implant-abutment screws in three different implant systems under simulated occlusal loads. Materials and methods: Three abutments to implant internal hex joint systems were simulated by using the 3-dimensional finite element analysis; (1) Nobel Biocare replace tapered (2) Uniti (3) Lifecare self-threaded tapered ceme...

متن کامل

Influence of abutment materials on the implant-abutment joint stability in internal conical connection type implant systems

PURPOSE This study evaluated the influence of abutment materials on the stability of the implant-abutment joint in internal conical connection type implant systems. MATERIALS AND METHODS Internal conical connection type implants, cement-retained abutments, and tungsten carbide-coated abutment screws were used. The abutments were fabricated with commercially pure grade 3 titanium (group T3), c...

متن کامل

Abutment fracture and the technique Used to Remove the fracture portion of the Abutmet: Case Report

Abstract This clinical report presents a case of abutment fracture, due to occlusal overload, occurred with an implant-supported restoration on right mandibular first molar of a female patient. After 6 months of prosthetic reconstruction implants, the patient returned to the clinic due to the loss of implant crown. The fractured fragment of the abutment with the screw protec...

متن کامل

Preload and Torque Removal Evaluation of Three Different Abutment Screws for Single Standing Implant Restorations

Several authors still consider the mechanical problems of fracture and component loosening as the main causes of failure of implant-supported restorations. The purpose of this in vitro study was to compare the preload of three types of screw for transmucosal abutment attachment used in single implant-supported prosthesis through strain gauge and removal torque measurements. Three external hex f...

متن کامل

Measuring abutment/implant joint integrity with the Periotest instrument.

Maintenance of the integrity of the abutment/implant interface is essential and is dependent on the abutment screw retaining a preload. Evaluation of this joint is usually done by manual assessment. The purpose of the current study was to determine whether the Periotest instrument could be used to evaluate abutment screw loosening. A custom-designed apparatus was constructed to measure abutment...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The International journal of oral & maxillofacial implants

دوره 16 3  شماره 

صفحات  -

تاریخ انتشار 2001