Dosimetric impact of source-positioning uncertainty in high-dose-rate balloon brachytherapy of breast cancer
نویسنده
چکیده
PURPOSE To evaluate the dosimetric impact of source-positioning uncertainty in high-dose-rate (HDR) balloon brachytherapy of breast cancer. MATERIAL AND METHODS For 49 HDR balloon patients, each dwell position of catheter(s) was manually shifted distally (+) and proximally (-) with a magnitude from 1 to 4 mm. Total 392 plans were retrospectively generated and compared to corresponding clinical plans using 7 dosimetric parameters: dose (D95) to 95% of planning target volume for evaluation (PTV_EVAL), and volume covered by 100% and 90% of the prescribed dose (PD) (V100 and V90); skin and rib maximum point dose (Dmax); normal breast tissue volume receiving 150% and 200% of PD (V150 and V200). RESULTS PTV_EVAL dosimetry deteriorated with larger average/maximum reduction (from ± 1 mm to ± 4 mm) for larger source position uncertainty (p value < 0.0001): from 1.0%/2.5%, 3.3%/5.9%, 6.3%/10.0% to 9.8%/14.5% for D95; from 1.0%/2.6%, 3.1%/5.7%, 5.8%/8.9% to 8.7%/12.3% for V100; from 0.2%/1.5%, 1.0%/4.0%, 2.7%/6.8% to 5.1%/10.3% for V90. ≥ ± 3 mm shift reduced average D95 to < 95% and average V100 to < 90%. While skin and rib Dmax change was case-specific, its absolute change (∣Δ(Value)∣) showed that larger shift and high dose group had larger variation compared to smaller and lower dose group (p value < 0.0001), respectively. Normal breast tissue V150 variation was case-specific and small. Average ∣Δ(V150)∣ was 0.2 cc for the largest shift (± 4 mm) with maximum < 1.7 cc. V200 was increased with higher elevation for larger shift: from 6.4 cc/9.8 cc, 7.0 cc/10.1 cc, 8.0 cc/11.3 cc to 9.2 cc/ 13.0 cc. CONCLUSIONS The tolerance of ± 2 mm recommended by AAPM TG 56 is clinically acceptable in most clinical cases. However, special attention should be paid to a case where both skin and rib are located proximally to balloon, and the orientation of balloon catheter(s) is vertical to these critical structures. In this case, sufficient dosimetric planning margins are required.
منابع مشابه
Calculation of absorbed dose rate in the lungs, ribs and skin by the mammosite applicator in breast cancer brachytherapy with MCNP
Introduction: Breast cancer is the most common cancer among women. Brachytherapy is one of the ways to treat breast cancer. In brachytherapy treatments, high-dose sources are used in interstitial placement. Here, Iridium-192 is used. This source emits gamma rays with energies ranging from 136 kV to 1060 kV electron volts. The activity of the used spring is 10 Ci. The simulated ...
متن کاملDosimetric characterization of a high dose rate 192I source for brachytherapy application using Monte Carlo simulation and benchmarking with thermoluminescent dosimetry
Background: The purpose of this project was to derive the brachytherapy dosimetric functions described by American Association of Physicists in Medicine (AAPM) TG-43 U1 based on high dose rate 192I sources. Materials and Methods: The method utilized included both simulation of the designed Polymethyl methacrylate (PMMA) phantom using the Monte Carlo of MCNP4C and benchmarking of the simulation ...
متن کاملEstimation of dosimetric parameters of I-125 brachytherapy source model 6711 using GATE8.1 code
Brachytherapy is one type of internal radiation therapy where radiation sources, which are usually encapsulated are placed as close as possible to the tumor site inside the patient's body. In this technique, it is important to determine dose distribution around the brachytherapy capsule. Hereby, in this paper, dosimetric parameters of I-125 brachytherapy source model 6711 are estimated accordin...
متن کاملProof of Symmetric Dose Distribution in Mammosite Applicator in Breast Brachytherapy with MCNP Simulation
Introduction: In brachytherapy treatments, high-dose sources are used in interstitial placement. In this paper, the modeling of a common breast cancer treatment applicator called MammoSite is done by the Monte Carlo simulation code. Then, to study and calculate the amount of doses in the breast and the dose rate of the organs at risk of radiation, including: lungs, ribs and ski...
متن کاملDetermination of TG-43 Dosimetric Parameters for Photon Emitting Brachytherapy Sources
Objective: Brachytherapy sources are widely used for the treatment of cancer. The report of Task Group No. 43 (TG-43) of American Association of Physicists in Medicine is known as the most common method for the determination of dosimetric parameters for brachytherapy sources. The aim of this study is to obtain TG-43 dosimetric parameters for 60Co, 137Cs, 192Ir and 103Pd brachyt...