Calibration of an ingestible temperature sensor.

نویسندگان

  • A P Hunt
  • I B Stewart
چکیده

An ingestible telemetric sensor for measuring core body temperature is increasingly being utilized in occupational and athletic studies of heat strain. There is a need for a uniform method of calibrating these sensors in the scientific community in order to effectively compare the results of different researchers. The purpose of the present investigation was to determine and present such a calibration procedure. Sensors were placed in a water bath heated to nine discrete temperatures, and the recorded values were compared to that of a traceable thermometer. It was observed that sensor 2 recorded temperatures higher than sensors 1 and 3, and that all sensors were higher than the traceable thermometer, highlighting the need for a calibration procedure. The findings of this study suggest a number of recommendations for a calibration procedure including: (1) four water bath temperatures in the range of 33-41 degrees C should be utilized; (2) sensors should be immersed for a minimum of 4 min prior to taking a measurement; (3) a linear regression relating sensor temperature to a traceable thermometer is an appropriate method to adjust raw data. Switching the sensor off after calibration and reactivating it prior to ingestion will not influence the accuracy of temperature measurement.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The Systematic Bias of Ingestible Core Temperature Sensors Requires a Correction by Linear Regression

An accurate measure of core body temperature is critical for monitoring individuals, groups and teams undertaking physical activity in situations of high heat stress or prolonged cold exposure. This study examined the range in systematic bias of ingestible temperature sensors compared to a certified and traceable reference thermometer. A total of 119 ingestible temperature sensors were immersed...

متن کامل

Validation of an ingestible temperature data logging and telemetry system during exercise in the heat

Aim: Intestinal temperature telemetry systems are promising monitoring and research tools in athletes. However, the additional equipment that must be carried to continuously record temperature data limits their use to training. The purpose of this study was to assess the validity and reliability of a new gastrointestinal temperature data logging and telemetry system (e-Celsius™) during water ba...

متن کامل

A New Approach to Self-Localization for Mobile Robots Using Sensor Data Fusion

This paper proposes a new approach for calibration of dead reckoning process. Using the well-known UMBmark (University of Michigan Benchmark) is not sufficient for a desirable calibration of dead reckoning. Besides, existing calibration methods usually require explicit measurement of actual motion of the robot. Some recent methods use the smart encoder trailer or long range finder sensors such ...

متن کامل

Using an Ingestible Telemetric Temperature Pill to Assess Gastrointestinal Temperature During Exercise.

Exercise results in an increase in core body temperature (Tc), which may reduce exercise performance and eventually can lead to the development of heat-related disorders. Therefore, accurate measurement of Tc during exercise is of great importance, especially in athletes who have to perform in challenging ambient conditions. In the current literature a number of methods have been described to m...

متن کامل

Real Time Calibration of Strap-down Three-Axis-Magnetometer for Attitude Estimation

Three-axis-magnetometers (TAMs) are widely utilized as a key component of attitude determination subsystems and as such are considered the corner stone of navigation for low Earth orbiting (LEO) space systems. Precise geomagnetic-based navigation demands accurate calibration of the magnetometers. In this regard, a complete online calibration process of TAM is developed in the current research t...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Physiological measurement

دوره 29 11  شماره 

صفحات  -

تاریخ انتشار 2008