Evaluating and defining the sharpness of intraocular lenses: microedge structure of commercially available square-edged hydrophilic intraocular lenses.

نویسندگان

  • Liliana Werner
  • Manfred Tetz
  • Ines Feldmann
  • Michael Bücker
چکیده

PURPOSE To evaluate the microstructure of the edges of currently available hydrophilic acrylic intraocular lenses (IOLs) in terms of their deviation from an "ideal" square as a follow-up of preliminary in vitro studies of experimental poly(methyl methacrylate) IOLs and commercially available foldable hydrophobic IOLs. SETTING Berlin Eye Research Institute, Berlin, Germany. METHODS Twenty-four designs of hydrophilic acrylic IOLs were used in this study. For each design, a +20.0 diopter (D) IOL and a +0.0 D IOL (or the lowest available plus dioptric power) were evaluated. The IOL edge was imaged under low-vacuum (0.7 torr), high-magnification scanning electron microscopy (SEM) using an environmental microscope and standardized technique. The photographs were imported to a digital computer program, and the area above the posterior-lateral edge, representing the deviation from a perfect square, was measured in square microns. RESULTS Currently available hydrophilic acrylic IOLs labeled as square edged had an area of deviation from a perfect square ranging from 60.84 to 871.51 microm(2) for the +20.0 D IOLs and from 35.52 to 826.55 microm(2) for the low-diopter IOLs. Although some differences in edge finishing between the IOLs analyzed were observed, edge surfaces of hydrophilic acrylic IOLs appeared overall smooth under environmental SEM. CONCLUSIONS Analysis of the microstructure of the optic edge of currently available square-edged hydrophilic acrylic IOLs showed a large variation of the deviation area from a perfect square.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Anterior Segment Intraocular Lens Edge Profiles of Currently Available Intraocular Lenses and Recent Improvements

We evaluated the microstructure of the edges of currently available foldable intraocular lenses (IOLs). The methodology used assisted an IOL manufacturer (Hoya, Japan) to optimise the edge profile of its lenses. Ten designs of hydrophobic acrylic, six designs of silicone and 24 designs of hydrophilic acrylic IOLs were imaged under scanning electron microscopy (SEM), and the photographs were imp...

متن کامل

Investigation of Intraocular Lenses Surface Roughness by Atomic Force Microscopy

The aim of this study was to evaluate the surface roughness of intraocular lenses (IOLs) generated by the manufacturing process and to determine the roughness parameters of 3D surface using atomic force microscopy (AFM). Intraocular lenses commercially available from a single manufacturer: Pharmacia & Upjohn Co. were investigated. Three intraocular lenses, model 911A CeeOn Edge, were analyzed o...

متن کامل

Findings of an experimental study in a rabbit model on posterior capsule opacification after implantation of hydrophobic acrylic and hydrophilic acrylic intraocular lenses

PURPOSE Study on cell growth on the posterior capsule after implantation of hydrophobic acrylic (Acrysof SA 60 AT) and hydrophilic acrylic (Akreos Disc) intraocular lenses (IOL) in a rabbit model and comparison of posterior capsule opacification (PCO). METHODS Phacoemulsification was performed in 22 rabbit eyes, and two different IOL types (Acrysof SA60 AT and Akreos Disc) were implanted. The...

متن کامل

Phakic Intraocular Lenses and their Special Indications

Phakic intraocular lenses revolutionize refractive surgery and continue to serve as an excellent option for vision correction in patients who are not ideal candidates for laser vision correction. This article will review special indications of phakic intraocular lenses in the clinical practice.

متن کامل

Comparison of hydrophobic and hydrophilic intraocular lens in preventing posterior capsule opacification after cataract surgery

BACKGROUND Posterior capsular opacification (PCO) is a common long-term complication of cataract surgery. Intraocular lens design and material have been implicated in influencing the development of PCO. This study evaluated the association of hydrophobic and hydrophilic intraocular lenses on preventing PCO. METHODS Medline, Cochrane, EMBASE, and Google Scholar databases were searched until Au...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of cataract and refractive surgery

دوره 35 3  شماره 

صفحات  -

تاریخ انتشار 2009