Miniaturized FDDA and CMOS Based Potentiostat for Bio-Applications
نویسندگان
چکیده
A novel fully differential difference CMOS potentiostat suitable for neurotransmitter sensing is presented. The described architecture relies on a fully differential difference amplifier (FDDA) circuit to detect a wide range of reduction-oxidation currents, while exhibiting low-power consumption and low-noise operation. This is made possible thanks to the fully differential feature of the FDDA, which allows to increase the source voltage swing without the need for additional dedicated circuitry. The FDDA also reduces the number of amplifiers and passive elements in the potentiostat design, which lowers the overall power consumption and noise. The proposed potentiostat was fabricated in 0.18 µm CMOS, with 1.8 V supply voltage. The device achieved 5 µA sensitivity and 0.99 linearity. The input-referred noise was 6.9 µV rms and the flicker noise was negligible. The total power consumption was under 55 µW. The complete system was assembled on a 20 mm × 20 mm platform that includes the potentiostat chip, the electrode terminals and an instrumentation amplifier for redox current buffering, once converted to a voltage by a series resistor. the chip dimensions were 1 mm × 0.5 mm and the other PCB components were off-chip resistors, capacitors and amplifiers for data acquisition. The system was successfully tested with ferricyanide, a stable electroactive compound, and validated with dopamine, a popular neurotransmitter.
منابع مشابه
A Miniaturized Cmos Microelectrode Array System for Single Droplet Electrochemistry Applications
This paper presents an individually addressable 32x32 microelectrode array (MEA) system with built-in CMOS potentiostat for analyzing single droplets and ultra-small volume samples. The proposed system has advantages of smaller sample requirement, higher density, small form factor, low power and higher sensitivity and advanced functionality over the conventional electrode array. Each electrode ...
متن کاملA miniaturized low-power wireless remote environmental monitoring system based on electrochemical analysis
In this paper, we report a miniaturized low-power wireless remote environmental monitoring system. This system has been developed for on-site monitoring of water pollution by heavy-metal ions. The system is composed of three parts: an electrochemical sensor module using microfabricated electrodes for detecting heavy-metal contamination in sample water; a custom potentiostat module including rea...
متن کاملLow-Cost Impedance Measurements for Lab-on-a-Chip Architectures: Towards Potentiostat Miniaturization
The development of miniaturized potentiostats capable of measuring in a wide range of conditions and with full characteristics (e.g., wide bandwidth and capacitive/inductive contribution to sensor’s impedance) is still an unresolved challenge in bioelectronics. We present a simple analogue design coupled to a digital filter based on a lock-in amplifier as an alternative to complex architectures...
متن کاملElectrochemical Microsensors for the Detection of Cadmium(II) and Lead(II) Ions in Plants
Routine determination of trace metals in complex media is still a difficult task for many analytical instruments. The aim of this work was to compare three electro-chemical instruments [a standard potentiostat (Autolab), a commercially available miniaturized potentiostat (PalmSens) and a homemade micropotentiostat] for easy-to-use and sensitive determination of cadmium(II) and lead(II) ions. Th...
متن کاملWide Dynamic Range CMOS Potentiostat for Amperometric Chemical Sensor
Presented is a single-ended potentiostat topology with a new interface connection between sensor electrodes and potentiostat circuit to avoid deviation of cell voltage and linearly convert the cell current into voltage signal. Additionally, due to the increased harmonic distortion quantity when detecting low-level sensor current, the performance of potentiostat linearity which causes the detect...