Quantification and reduction of peripheral dose from leakage radiation on Siemens primus accelerators in electron therapy mode

نویسندگان

  • Collins Yeboah
  • Alex Karotki
  • Dylan Hunt
  • Rick Holly
چکیده

In this work, leakage radiation from EA200 series electron applicators on Siemens Primus accelerators is quantified, and its penetration ability in water and/or the shielding material Xenolite-NL established. Initially, measurement of leakage from 10 x 10 - 25 x 25 cm2 applicators was performed as a function of height along applicator and of lateral distance from applicator body. Relative to central-axis ionization maximum in solid water, the maximum leakage in air observed with a cylindrical ion chamber with 1 cm solid water buildup cap at a lateral distance of 2 cm from the front and right sidewalls of applicators were 17% and 14%, respectively; these maxima were recorded for 18 MeV electron beams and applicator sizes of >or=20 x 20 cm2. In the patient plane, the applicator leakage gave rise to a broad peripheral dose off-axis distance peak that shifted closer to the field edge as the electron energy increases. The maximum peripheral dose from normally incident primary electron beams at a depth of 1 cm in a water phantom was observed to be equal to 5% of the central-axis dose maximum and as high as 9% for obliquely incident beams with angles of obliquity <or=40 degrees . Measured depth-peripheral dose curves showed that the "practical range" of the leakage electrons in water varies from approximately 1.4 to 5.7 cm as the primary electron beam energy is raised from 6 to 18 MeV. Next, transmission measurements of leakage radiation through the shielding material Xenolite-NL showed a 4 mm thick sheet of this material is required to attenuate the leakage from 9 MeV beams by two-thirds, and that for every additional 3 MeV increase in the primary electron beam energy, an additional Xenolite-NL thickness of roughly 2 mm is needed to achieve the aforementioned attenuation level. Finally, attachment of a 1 mm thick sheet of lead to the outer surface of applicator sidewalls resulted in a reduction of the peripheral dose by up to 80% and 74% for 9 and 18MeV beams, respectively. This sidewall modification had an insignificant effect on the clinical depth dose, cross-axis beam profiles, and output factors.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Benchmarking of Siemens Linac in Electron Modes: 8-14 MeV Electron Beams

Introduction: Radiation therapy using electron beams is a promising method due to its physical dose distribution. Monte Carlo (MC) code is the best and most accurate technique for forespeaking the distribution of dose in radiation treatment of patients.Materials and Methods: We report an MC simulation of a linac head and depth dose on central axis, along with profile calculations. The purpose o...

متن کامل

Siemens primus accelerator simulation using EGSnrc Monte Carlo code and gel dosimetry validation with optical computed tomography system by EGSnrc code

Monte Carlo method is the most accurate method for simulation of radiation therapy equipment. The linear accelerators (linac) are currently the most widely used machines in radiation therapy centers. Monte Carlo modeling of the Siemens Primus linear accelerator in 6 MeV beams was used. Square field size of 10 × 10 cm2 produced by the jaws was compared with TLD. Head simulation of Siemens accele...

متن کامل

The Quality Control of Intensity Modulated Radiation Therapy (IMRT) for ONCOR Siemens Linear Accelerators Using Film Dosimetry

Introduction Intensity Modulated Radiation Therapy (IMRT) has made a significant progress in radiation therapy centers in recent years. In this method, each radiation beam is divided into many subfields that create a field with a modulated intensity. Considering the complexity of this method, the quality control for IMRT is a topic of interest for researchers. This article is about the various ...

متن کامل

Assessment of secondary neutron dose due to dental restorations in head and neck radiation therapy

Introduction: One of scientific concern is increasing of unwanted neutron dose to the patient, in head and neck radiation therapy due to the presence of some isotopes in dental restorations and head of medical linac. The aim of this study is to measure the equivalent dose of thermal and fast neutron due to head of Siemens Primus Linac and a healthy tooth, Amalgam, Ni-Cr alloy a...

متن کامل

Dental Materials Effect in Neutron Contamination: Electron Mode of a Linac

Background: Neutron contamination is produced in electron beams of linac when tooth or dental materials are located in the path of beam. Objective: This study aims to determine the neutron dose contamination from different dental restoration materials in electron mode of a linac. Material and Methods: In this experimental study, the neutron dose contamination was calculated in the p...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 11  شماره 

صفحات  -

تاریخ انتشار 2010