Dose perturbation study in a multichannel breast brachytherapy device

نویسندگان

  • Eric D. Slessinger
  • Rodney Fletcher
  • Indra J Das
چکیده

PURPOSE A study was conducted to determine the dosimetric effects resulting from air pockets and high atomic number (Z) contrast medium within a multichannel breast brachytherapy device. MATERIAL AND METHODS A 5-6 cm diameter Contura (SenoRx) brachytherapy device was inflated using 37 cm(3) of saline. Baseline dose falloff from an HDR Iridium-192 source was measured with the Iridium source centered in the central channel and an anterior off-center channel. Data were collected at distances from 1 to 50 mm. Comparison studies were conducted with identically inflated volume containing varied air pocket volumes (1-4 cm(3)) and concentrations of contrast solution (3%, 6%, and 9% by volume). Dose perturbation factors (DPF) were computed and evaluated. RESULTS Dose perturbations due to air pockets and contrast solutions were observed. As the volume of air increased, the DPF increased by approximately 2.25%/cm(3). The effect was consistent for both channels. The contrast effects were more complex. The 3% contrast media had minimal dose perturbation. The 6% contrast solution caused dose reduction of 1.0% from the central channel but 1.5% dose increase from the anterior channel. The 9% contrast solution caused dose reductions by 4.0% (from central channel) and 3.0% (from anterior channel). The DPF from all contrast solutions moderated with increasing distance. CONCLUSIONS Dose perturbations due to air pockets and high-Z contrast solution can be significant. It is important to control these effects to avoid dose errors.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Monte Carlo investigation on precise dosimetry of HDR breast brachytherapy with Accuboost

Introduction: Accuboost is a HDR brachytherapy system in early stages breast cancer treatment. This device provides a completely noninvasive procedure with parallel-opposed radiation from two immobilizing peripheral applicators that caused it a preferred option of modalities to choose. In most commercial treatment planning systems, tissues are considered as a simple water phan...

متن کامل

The Effect of Breast Phantom Material on the Dose Distribution in AccuBoost Brachytherapy

Introduction: Long-term teletherapy program is not suitable for old and working patients and those living in areas where little access to primary health care is available. Accelerated partial breast irradiation (APBI) using high dose rate (HDR) brachytherapy is an appropriate alternative for these patients due to its limited number of fractions. The AccuBoost is a system for delivering APBI. Th...

متن کامل

Calculation of absorbed dose rate in the lungs, ribs and skin by the mammosite applicator in breast cancer brachytherapy with MCNP

Introduction: Breast cancer is the most common cancer among women. Brachytherapy is one of the ways to treat breast cancer. In brachytherapy treatments, high-dose sources are used in interstitial placement. Here, Iridium-192 is used. This source emits gamma rays with energies ranging from 136 kV to 1060 kV electron volts. The activity of the used spring is 10 Ci. The simulated ...

متن کامل

Proof of Symmetric Dose Distribution in Mammosite Applicator in Breast Brachytherapy with MCNP Simulation

Introduction: In brachytherapy treatments, high-dose sources are used in interstitial placement. In this paper, the modeling of a common breast cancer treatment applicator called MammoSite is done by the Monte Carlo simulation code. Then, to study and calculate the amount of doses in the breast and the dose rate of the organs at risk of radiation, including: lungs, ribs and ski...

متن کامل

Consideration of accurate dose distribution for Contura balloon catheter in breast brachytherapy with MCNP5 simulation

Introduction: Contura is a balloon applicator that used to deliver intra cavity radiation for Partial Breast Irradiation (PBI) treatments. Contura with 4 peripheral lumens surrounding the central, for loading HDR sources from an after loader system can provide an asymmetric dose distribution and gives a more flexible shape of dose than earlier balloon applicators such as Mammo...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 3  شماره 

صفحات  -

تاریخ انتشار 2011