Creatinine measurement proficiency testing: assignment of matrix-adjusted ID GC-MS target values.
نویسندگان
چکیده
The results of an external quality-assessment experiment for serum creatinine measurement are described. Fifty-one laboratories performed quintuplicate analyses during three different analytical runs on six lyophilized sera and two frozen human serum pools. Isotope dilution gas chromatography-mass spectrometry (ID GC-MS) target values were assigned to all the materials. Intralaboratory within- and between-run imprecision results were very similar for all the materials tested (CV < or = 2.20% and < or = 4.70%, respectively). The overall imprecision obtained was high (CV 6.5-20.0%) because of increased interlaboratory-intermethod variability. A significant positive bias (+ 9.2-+43.7%) was found for all the materials at lower creatinine concentration. By using two human sera at different concentrations, we could calculate the constant and the proportional calibration bias displayed by each peer group. The majority of the lyophilized materials showed a behavior divergent from the frozen pools, indicating matrix-related problems. We propose a new algorithm for calculating matrix bias correction factor instrument-reagent specific for each material.
منابع مشابه
Routine serum creatinine measurements: how well do we perform?
BACKGROUND The first aim of the study was to investigate the accuracy and intra-laboratory variation of serum creatinine measurements in clinical laboratories in Flanders. The second purpose was to check the effect of this variation in serum creatinine concentration results on the calculated estimated glomerular filtration rate (eGFR) and the impact on classification of patients into a chronic ...
متن کاملReference measurement procedure for total glycerides by isotope dilution GC-MS.
BACKGROUND The CDC's Lipid Standardization Program established the chromotropic acid (CA) reference measurement procedure (RMP) as the accuracy base for standardization and metrological traceability for triglyceride testing. The CA RMP has several disadvantages, including lack of ruggedness. It uses obsolete instrumentation and hazardous reagents. To overcome these problems the CDC developed an...
متن کاملCertification of a new selenized yeast reference material (SELM-1) for methionine, selenomethinone and total selenium content and its use in an intercomparison exercise for quantifying these analytes.
A new selenized yeast reference material (SELM-1) produced by the Institute for National Measurement Standards, National Research Council of Canada (INMS, NRC) certified for total selenium (2,059+/-64 mg kg(-1)), methionine (Met, 5,758+/-277 mg kg(-1)) and selenomethionine (SeMet, 3,431+/-157 mg kg(-1)) content is described. The +/-value represents an expanded uncertainty with a coverage factor...
متن کاملDetermination of creatinine in human serum by isotope dilution-mass spectrometry. Definitive methods in clinical chemistry, IV.
A method for the determination of creatinine in human serum by isotope dilution-mass spectrometry is described. The analytical procedure comprises the following steps: Addition of [13C,15N2 )creatinine to the serum sample; ion exchange chromatography on the cation exchange resin AG 50W-X2; formation of the trimethylsilyl derivative; gas liquid chromatography-mass spectrometry (GC-MS); selected ...
متن کاملState-of-the-art of serum testosterone measurement by isotope dilution-liquid chromatography-tandem mass spectrometry.
BACKGROUND The recent interest of clinical laboratories in developing serum testosterone assays based on isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS) stems from the lack of accuracy of direct immunoassays. In this study, we assessed the accuracy and state of standardization (traceability) of 4 published ID-LC-MS/MS procedures in a method comparison with an ID-ga...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinical chemistry
دوره 43 8 Pt 1 شماره
صفحات -
تاریخ انتشار 1997