EyeSys corneal topography measurement applied to calibrated ellipsoidal convex surfaces.

نویسنده

  • W A Douthwaite
چکیده

AIMS/BACKGROUND This study was carried out to assess the accuracy of the EyeSys videokeratoscope by using convex ellipsoidal surfaces of known form. METHODS PMMA convex ellipsoidal buttons were calibrated using Form Talysurf analysis which allowed subsequent calculation of the vertex radius and p value of the surface. The EyeSys videokeratoscope was used to examine the same ellipsoids. The tabular data provided by the instrument software were used to plot a graph of r2 versus y2 where r is the measured radius at y, the distance from the corneal point being measured to the surface vertex. The intercept on the ordinate of this graph gives the vertex radius and the slope the p value. The results arising from the Talysurf and the EyeSys techniques were compared. RESULTS The EyeSys videokeratoscope gave readings for both vertex radius and p value that were higher than those of the Talysurf analysis. The vertex radius was around 0.1 mm greater. The p value results were similar by the two methods for p values around unity but the EyeSys results were higher and the discrepancy increased as the p value approached that of a paraboloid. CONCLUSIONS Although the videokeratoscope may be useful in comparative studies of the cornea, there must be some doubt about the absolute values displayed. The disagreement is sufficiently large to suggest that the instrument may not be accurate enough for contact lens fitting purposes.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of videokeratoscope and autokeratometer measurements on ellipsoid surfaces and human corneas.

BACKGROUND Various studies have compared the accuracy and repeatability of autokeratometers and videokeratoscopes using calibrated convex surfaces. We investigate the agreement between the Topcon KR-3500 autokeratometer and the EyeSys videokeratoscope on human corneas and calibrated convex surfaces. METHODS Measurements were obtained from 30 convex ellipsoidal buttons and 20 right eyes of 20 ...

متن کامل

Corneal topography by keratometry.

AIMS To investigate the ability of a telecentric keratometer to describe the asphericity and curvature of convex ellipsoidal surfaces and human corneas. METHODS 22 conicoidal convex surfaces and 30 human corneas were examined by conventional keratometry. Additional keratometric measurements were made when the surface was tilted in the horizontal plane relative to the instrument optical axis. ...

متن کامل

[Diurnal variations of central and paracentral corneal thickness and curvature].

OBJECTIVE Corneal topography and thickness has clinical importance in contact lens fitting and refractive surgery, however repeated measurement of corneal thickness and curvature is complicated by the natural phenomenon of diurnal variation. Our aim was determine diurnal variations of central and paracentral corneal thickness and curvature over a period of ten hours. MATERIALS AND METHODS Cor...

متن کامل

Comparison of Topographic Technologies in Anterior Surface Mapping of Keratoconus using Two Display Algorithms and Six Corneal Topography Devices

Purpose: To evaluate anterior surface topographic technologies and display algorithms in mapping keratoconus. Materials and methods: A total of 27 eyes of 17 subjects clinically diagnosed with keratoconus were imaged on six topographers: EyeSys, Alcon EyeMap, Keratron, TMS-1, Orbscan and PAR corneal topography system. Axial distance (AD) and instantaneous radius of curvature (IROC) algorithms w...

متن کامل

Extent and effect of surface tilt on the data display of the EyeSys videokeratoscope.

AIMS This study examined the occurrence, magnitude, and the consequences of a possible tilt between the corneal surface and optical axis of the EyeSys videokeratoscope. METHODS Initially, a theoretical model was developed to calculate the angle of tilt. The predictions of the model were verified empirically using a convex conicoid surface and were found to predict the tilt to within 0.5 degre...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The British journal of ophthalmology

دوره 79 9  شماره 

صفحات  -

تاریخ انتشار 1995