Examination of steroid release from screw-in pacing leads.
نویسندگان
چکیده
Although constant progress in pacing lead design and life time of pacemaker battery has contributed to the clinical success of this treatment option, the mechanism of drug release from steroid-eluting pacing leads is not completely understood. Besides, drug concentrations of steroids released into the heart-tissue remain unclear. Therefore, three different dissolution test methods were examined: a non-standardized (reagent tube), a compendial USP (paddle) and a hydrogel method. The experimental results indicate that dissolution from pacing leads is governed by a diffusion-controlled mechanism and occurs over a prolonged time. Furthermore, the integration of a hydrogel compartment seemed to have an impact on drug release leading to a decrease in release rate. Nevertheless, the experimental results emphasize different experimental parameters in dissolution testing may have a distinct influence on drug dissolution.
منابع مشابه
Acute performance of steroid-eluting screw-in leads for atrial free wall pacing.
The aim of this study was to clarify the acute performance of steroid-eluting screw-in leads in comparison with that of nonsteroid screw-in leads for atrial free wall pacing. In 114 cases (68 males, 46 females, average age 70 years) with atrial free wall pacing by screw-in leads, pacing thresholds and P-wave amplitudes were compared at the time of implantation and 1 week later between 68 cases ...
متن کاملPacing therapy in children.
BACKGROUND The importance of ventricular pacing site in pediatric pacemaker therapy has gradually become recognized. We reviewed our experience with a left ventricular (LV)-prioritized pacing strategy. METHODS AND RESULTS: Between 2000 and 2012, 60 patients underwent 76 permanent pacemaker implantations. Eight of the 29 reoperations involved ventricular lead repositioning for pacing-induced ven...
متن کاملEntire Removal of Screw-In Pacing Leads 3 Years after Implantation
OBJECTIVE To report that screw-in type pacing leads can be removed by screw retraction even after a significant anchoring period. PATIENT A 78-year-old woman who visited our hospital for skin erosion over a pacemaker that had been implanted 3 years previously and had migrated from the subclavicular area to the axilla. METHODS Culture revealed a local staphylococcus infection. We placed a ne...
متن کاملEpicardial left ventricular leads via minimally invasive technique: a role of steroid eluting leads
BACKGROUND We retrospectively assessed two types of sutureless screw-in left ventricular (LV) leads (steroid eluting vs. non-steroid eluting) in cardiac resynchronization therapy (CRT) implantation with regards to their electrical performance. METHODS Between March 2008 and May 2014 an epicardial LV lead was implanted in 32 patients after failed transvenous LV lead placement using a left-side...
متن کاملAcute changes in pacing threshold and R- or P-wave amplitude during permanent pacemaker implantation.
This study examines the changes in pacing threshold and R- or P-wave amplitude during the first 30 minutes after implantation of tined and screw-in leads. The leads examined were those of 1 manufacturer (Medtronic) and consisted of 3 ventricular pacing leads (model numbers 6957 unipolar screw-in [11 patients], 6961 unipolar tined [12 patients] and 6962 bipolar tined [7 patients]) and 1 atrial l...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Biomedizinische Technik. Biomedical engineering
دوره 57 Suppl 1 شماره
صفحات -
تاریخ انتشار 2012