MR imaging-related heating of deep brain stimulation electrodes: in vitro study.
نویسندگان
چکیده
BACKGROUND AND PURPOSE Recent work has shown a potential for excessive heating of deep brain stimulation electrodes during MR imaging. This in vitro study investigates the relationship between electrode heating and the specific absorption rate (SAR) of several MR images. METHODS In vitro testing was performed by using a 1.5-T MR imaging system and a head transmit-receive coil, with bilateral deep brain stimulation systems positioned in a gel saline-filled phantom, and temperature monitoring with a fluoroptic thermometry system. Standardized fast spin-echo sequences were performed over a range of high, medium, and low SAR values. Several additional, clinically important MR imaging techniques, including 3D magnetization prepared rapid acquisition gradient-echo imaging, echo-planar imaging, quantitative magnetization transfer imaging, and magnetization transfer-suppressed MR angiography, were also tested by using typical parameters. RESULTS A significant, highly linear relationship between SAR and electrode heating was found, with the temperature elevation being approximately 0.9 times the local SAR value. Minor temperature elevations, <1 degrees C, were found with the fast spin-echo, magnetization prepared rapid acquisition gradient-echo, and echo-planar clinical imaging sequences. The high dB/dt echo-planar imaging sequence had no significant heating independent of SAR considerations. Sequences with magnetization transfer pulses produced temperature elevations in the 1.0 to 2.0 degrees C range, which was less than theoretically predicted for the relatively high SAR values. CONCLUSION A potential exists for excessive MR imaging-related heating in patients with deep brain stimulation electrodes; however, the temperature increases are linearly related to SAR values. Clinical imaging sequences that are associated with tolerable temperature elevations in the <or=2.0 degrees C range at the electrode tips can be performed safely within an SAR range <2.4 W/kg local (0.9 W/kg whole body averaged).
منابع مشابه
Bilateral neurostimulation systems used for deep brain stimulation: in vitro study of MRI-related heating at 1.5 T and implications for clinical imaging of the brain.
Deep brain stimulation (DBS) is used increasingly in the field of movement disorders. The implanted electrodes create not only a prior risk to patient safety during MRI, but also a unique opportunity in the collection of functional MRI data conditioned by direct neural stimulation. We evaluated MRI-related heating for bilateral neurostimulation systems used for DBS with an emphasis on assessing...
متن کاملHeating induced near deep brain stimulation lead electrodes during magnetic resonance imaging with a 3 T transceive volume head coil.
Heating induced near deep brain stimulation (DBS) lead electrodes during magnetic resonance imaging with a 3 T transceive head coil was measured, modeled, and imaged in three cadaveric porcine heads (mean body weight = 85.47 ± 3.19 kg, mean head weight = 5.78 ± 0.32 kg). The effect of the placement of the extra-cranial portion of the DBS lead on the heating was investigated by looping the extra...
متن کاملRelationship of the P Angle with Stereotactic Arc in Intraoperative Outcome of Patients with Parkinson’s Disease Undergoing Deep Brain Stimulation
Background & Aim: DBS (deep brain stimulation) is a new and successful technique in treatment of symptoms of Parkinsonism especially after drug resistance. Research in this field is mostly designed for evolution of this technique. The present study aimed at evaluating the relationship between the angle formed in midsagittal and STN (sub-thalamic nucleus) axis line and recording l...
متن کاملNeurostimulation systems for deep brain stimulation: in vitro evaluation of magnetic resonance imaging-related heating at 1.5 tesla.
PURPOSE To assess magnetic resonance imaging (MRI)-related heating for a neurostimulation system (Activa Tremor Control System, Medtronic, Minneapolis, MN) used for chronic deep brain stimulation (DBS). MATERIALS AND METHODS Different configurations were evaluated for bilateral neurostimulators (Soletra Model 7426), extensions, and leads to assess worst-case and clinically relevant positionin...
متن کاملComments on MR heating tests of critical implants.
MAGNETIC RESONANCE IMAGING (MRI)-related heating for certain implants may be excessive, causing substantial patient injury. The determination of implant heating is particularly challenging because of the numerous variables that must be considered in order to properly identify both MRIand implant-related conditions that can impact the findings (1,2). Therefore, I applaud the recent commentary by...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- AJNR. American journal of neuroradiology
دوره 23 10 شماره
صفحات -
تاریخ انتشار 2002