Effects of slice orientation on reproducibility of sequential assessment of right ventricular volumes and ejection fraction: short-axis vs transverse SSFP cine cardiovascular magnetic resonance
نویسندگان
چکیده
BACKGROUND Test-retest reproducibility is of utmost importance in follow-up of right ventricular (RV) volumes and function; optimal slice orientation though is not yet known. We compared test-retest reproducibility and intra-/inter-observer variability of right ventricular (RV) volumes and function assessed with short-axis and transverse cardiovascular magnetic resonance (CMR). METHODS Eighteen volunteers underwent cine CMR for RV assessment obtaining ventricular coverage in short-axis and transverse slice orientation. Additional 2D phase contrast flow imaging of the main pulmonary artery (MPA) was performed. After complete repositioning repeat acquisitions were performed. Data sets were contoured by two blinded observers. Statistical analysis included Student's t-test, Bland-Altman plots, intra-class correlation coefficient (ICC) and 2-way ANOVA, SEM and minimal detectable difference calculations. RESULTS Heart rates (65.0 ± 7.4 vs. 67.6 ± 9.9 bpm; P = 0.1) and MPA flow (89.8 ± 16.6 vs. 87.2 ± 14.9 mL; P = 0.1) did not differ between imaging sessions. EDV and ESV demonstrated an inter-study bias of 0.4 %[-9.5 %,10.3 %] and 2.1 %[-12.3 %,16.4 %] for short-axis and 1.1 %[-7.3 %,9.4 %] and 0.8 %[-16.0 %,17.6 %] for transverse orientation, respectively. There was no significant interaction between imaging orientation and interstudy reproducibility (p = 0.395-0.824), intra-observer variability (p = 0.726-0.862) or inter-observer variability (p = 0.447-0.706) by 2-way ANOVA. Inter-observer agreement by ICC was greater for short axis versus transverse orientation for all parameters (0.769-0.986 vs. 0.625-0.983, respectively). Minimal detectable differences for short axis and transverse orientations were 10.1 mL/11.5 mL for EDV, 8.3 mL/8.4 mL for ESV and 4.1 % vs. 4.7 % for EF, respectively. CONCLUSION Short-axis and transverse orientation both provide reliable and reproducible measures for follow-up of RV volumes and global function. Therefore, additional transverse SSFP cine CMR may not necessarily be required if performed for the sole purpose of quantitative volumetric RV assessment.
منابع مشابه
Improved accuracy of quantitative assessment of left ventricular volume and ejection fraction by geometric models with steady-state free precession.
The purpose of this study was to determine whether steady-state free precession (SSFP) could improve accuracy of geometric models for evaluation of left ventricular (LV) function in comparison to turbo gradient echo (TGrE) and thereby reduce the acquisition and post-processing times, which are commonly long by use of the Simpson's Rule. In 25 subjects, cine loops of the complete heart in short ...
متن کاملEvaluation of left ventricular volumes and ejection fraction using fast steady-state cine MR imaging: comparison with left ventricular angiography.
Previous studies demonstrated that magnetic resonance (MR) imaging consistently underestimated angiographic measurements of left ventricular (LV) volumes. The purpose of this study was to determine whether MR imaging with steady-state free precession acquisition (SSFP) can provide improved accuracy and reproducibility in measuring cardiac function in comparison with fast spoiled gradient echo c...
متن کاملThe intra-observer reproducibility of cardiovascular magnetic resonance myocardial feature tracking strain assessment is independent of field strength
BACKGROUND Cardiovascular magnetic resonance myocardial feature tracking (CMR-FT) is a promising novel method for quantification of myocardial wall mechanics from standard steady-state free precession (SSFP) images. We sought to determine whether magnetic field strength affects the intra-observer reproducibility of CMR-FT strain analysis. METHODS We studied 2 groups, each consisting of 10 hea...
متن کاملComparison of long and short axis quantification of left ventricular volume parameters by cardiovascular magnetic resonance, with ex-vivo validation
BACKGROUND The purpose of the study was to compare the accuracy and evaluation time of quantifying left ventricular (LV), left atrial (LA) volume and LV mass using short axis (SAX) and long axis (LAX) methods when using cardiovascular magnetic resonance (CMR). MATERIALS AND METHODS We studied 12 explanted canine hearts and 46 patients referred for CMR (29 male, age 47 ± 18 years) in a clinica...
متن کاملCardiovascular magnetic resonance myocardial feature tracking detects quantitative wall motion during dobutamine stress
BACKGROUND Dobutamine stress cardiovascular magnetic resonance (DS-CMR) is an established tool to assess hibernating myocardium and ischemia. Analysis is typically based on visual assessment with considerable operator dependency. CMR myocardial feature tracking (CMR-FT) is a recently introduced technique for tissue voxel motion tracking on standard steady-state free precession (SSFP) images to ...
متن کامل