Introduction. To achieve a functional atraumatic cochlear implantation, intracochlear pressure changes during the procedure should be minimized. Postinsertional cable movements are assumed to induce intracochlear pressure changes. The aim of this study was to observe intracochlear pressure changes due to postinsertional cable movements. Materials and Methods. Intracochlear pressure changes were recorded in a cochlear model with a micro-pressure sensor positioned in the apical region of the cochlea model to follow the maximum amplitude and pressure gain velocity in intracochlear pressure. A temporal bone mastoid cavity was attached to the model to simulate cable positioning. The compared conditions were (1) touching the unsealed electrode, (2) touching the sealed electrode, (3) cable storage with an unfixed cable, and (4) cable storage with a fixed cable. Results. We found statistically significant differences in the occurrence of maximum amplitude and pressure gain velocity in intracochlear pressure changes under the compared conditions. Comparing the cable storage conditions, a cable fixed mode offers significantly lower maximum pressure amplitude and pressure gain velocity than the nonfixed mode. Conclusion. Postinsertional cable movement led to a significant pressure transfer into the cochlea. Before positioning the electrode cable in the mastoid cavity, fixation of the cable is recommended.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Cochlear implant electrode sealing techniques and related intracochlear pressure changes

BACKGROUND The inserted cochlear implanted electrode is covered at the site of the round window or cochleostomy to prevent infections and leakage. In a surgically hearing preservational concept, low intracochlear pressure changes are of high importance. The aim of this study was to observe intracochlear pressure changes due to different sealing techniques in a cochlear model. METHODS Cochlear...

متن کامل

Radiological and Histological Evaluation of Perimodular Electrodes: Impact on Cochlear Implant Development

Introduction: Different Perimodiolar electrodes were developed by many cochlear implant (CI) manufacturers in the past few years. Possible advantages are lower stimulation thresholds and a better channel separation. A position close to the modiolus can be facilitated by different means: preformed electrode shape (Nucleus ContourTM electrode, Cochlear Corp., Australia), additional positioning el...

متن کامل

An Intracochlear Pressure Sensor as a Microphone for a Fully Implantable Cochlear Implant.

OBJECTIVE To validate an intracochlear piezoelectric sensor for its ability to detect intracochlear pressure and function as a microphone for a fully implantable cochlear implant. METHODS A polyvinylidene fluoride (PVDF) piezoelectric pressure sensor was inserted into a human fresh cadaveric round window at varying depths. An external sound pressure stimulus was applied to the external audito...

متن کامل

Intracochlear Fluid Pressure Changes Related to the Insertional Speed of a CI Electrode

INTRODUCTION To preserve residual hearing the atraumaticity of the cochlea electrode insertion has become a focus of cochlear implant research. In addition to other factors, the speed of insertion is thought to be a contributing factor in the concept of atraumatic implantation. The aim of our study was to observe intracochlear fluid pressure changes due to different insertional speeds of an imp...

متن کامل

Evaluation of the Slope of Amplitude Growth Function Changes of the Electrically Evoked Action Potential in Three Months after Receiving the Device in Children with Cochlear Implant

Objective: In neural response telemetry, intracochlear electrodes stimulate the auditory nerve and record the neural responses. The electrical stimulation is sent to the auditory nerve by an electrode and the resulted response, called electrically evoked compound action potential, is recorded by an adjacent electrode. The most important clinical applications of this test are evaluation and moni...