Comparison of central corneal thickness estimated by an ultrasonic pachymeter and non-contact specular microscopy.

نویسندگان

  • Sadık Görkem Çevik
  • Rahmi Duman
  • Mediha Tok Çevik
  • Sertaç Argun Kıvanç
  • Berna Akova-Budak
  • Irfan Perente
  • Resat Duman
چکیده

Purpose: To compare central corneal thickness (CCT) measurements of healthy individuals obtained with ultrasonic pachymetry (UP) and non-contact specular microscopy (NCSM). Method: In total, 148 eyes of 74 subjects with no ocular or systemic diseases were included in the study. Central corneal thickness measurements of all patients performed with UP and NCCM were compared. Results: A total of 74 subjects (38 females) were included in this study. The mean age was 45.2 ± 18.4 (range 12-85) years. The mean central corneal thickness of all 148 eyes was 546.9 ± 40 μm with UP and 510.8 ± 42 μm with NCSM. The mean central corneal thickness measured with NCSM was 35 μm thinner than that measured with UP (p<0.001). A high degree of agreement was found between the two methods (r=0.942, p<0.001). Conclusions: Our results suggest that NCSM measures thinner corneas than UP and that the correction formula we identified should be applied when comparing between these two devices.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of corneal thickness measurements using Orbscan II, non-contact specular microscopy, and ultrasonic pachymetry in eyes after laser in situ keratomileusis.

AIMS To compare central corneal thickness measurements of three pachymetry devices in eyes after laser in situ keratomileusis (LASIK). METHODS Central corneal thickness was measured in 203 eyes after myopic LASIK. Orbscan II scanning slit topography (Bausch & Lomb), SP-2000P non-contact specular microscopy (Topcon), and ultrasonic pachymetry (Tomey) were used in this sequence. RESULTS Three...

متن کامل

SCIENTIFIC REPORT Comparison of corneal thickness measurements using Orbscan II, non-contact specular microscopy, and ultrasonic pachymetry in eyes after laser in situ keratomileusis

Aims: To compare central corneal thickness measurements of three pachymetry devices in eyes after laser in situ keratomileusis (LASIK). Methods: Central corneal thickness was measured in 203 eyes after myopic LASIK. Orbscan II scanning slit topography (Bausch & Lomb), SP-2000P non-contact specular microscopy (Topcon), and ultrasonic pachymetry (Tomey) were used in this sequence. Results: Three ...

متن کامل

Comparison of optical coherence reflectometry and ultrasound central corneal pachymetry.

In 50 eyes of 25 patients we prospectively measured the central corneal thickness (CCT) comparing the OLCR (Optical Low Coherence Reflectometry) pachymeter with the contact ultrasound pachymeter. The OLCR system was mounted on to a Haag-Streit slit lamp. Every single measurement was the result of 5 scans. With the contact ultrasound Sonomed pachymeter we performed 5 separate measurements and ca...

متن کامل

Comparison of Central Corneal Thickness Measurement Using Non-Contact and Contact Pachymetry Devices in Normal Eyes

Materials and Methods: This prospective, cross sectional comparative study evaluated 30 healthy subjects fulfilling the inclusion criteria reporting in AFIO Rawalpindi. CCT was measured in both eyes of subjects using noncontact specular microscope, Dual Scheimpflug analyzer and contact ultrasound pachymeter by a single investigator at the same time of the day. Pearson’s correlation coefficient ...

متن کامل

Comparison of central corneal thickness: ultrasound pachymetry versus slit-lamp optical coherence tomography, specular microscopy, and Orbscan

BACKGROUND Central corneal thickness (CCT) can be measured by using contact and non-contact methods. Ultrasound pachymetry (US pachymetry) is a contact method for measuring CCT and is perhaps the most commonly used method. However, non-contact methods like scanning slit topography (Orbscan II), slit-lamp optical coherence tomography (SL-OCT), and specular microscopy are also used. Not many stud...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Arquivos brasileiros de oftalmologia

دوره 79 5  شماره 

صفحات  -

تاریخ انتشار 2016