Multicentre ISI assignment and calibration of the INR measuring range of a new point-of-care system designed for home monitoring of oral anticoagulation therapy.

نویسندگان

  • Ingrid Leichsenring
  • Winfried Plesch
  • Volker Unkrig
  • Steve Kitchen
  • Dianne P Kitchen
  • Rhona Maclean
  • Bert Dikkeschei
  • Anton M H P van den Besselaar
چکیده

The new CoaguChek XS system is designed for use in patient selftesting. It is the successor of the current CoaguChek S system. The detection principle is based on the amperometric measurement of the thrombin activity initiated by starting the coagulation cascade using a human recombinant thromboplastin. This study was performed to assign the International SEnsitivity Index (ISI) to the new test according to the WHO guidelines for thromboplastins and plasmas used to control anticoagulant therapy, and to establish the measuring range of the new system. At four study sites a total of 90 samples of normal donors and 291 samples of warfarin-, phenprocoumon- or acenocoumarol-treated patients were included in the study. The ISI value of the new test was assigned against the human recombinant reference thromboplastin rTF/95 at each site using the samples from stabilized patients in the International Normalized Ratio (INR) range between 1.5 and 4.5 only. The new point-of-care system's measuring range between 0.8 and 8 INR was calibrated against the mean INR of rTF/95 and AD149 using polynomial regression. ISIs were (CV of the slope): Site 1: ISI 0.99 (1.1%); Site 2: ISI 1.02 (2.0%); Site 3: ISI 1.03 (1.1%); Site 4: ISI 1.00 (1.4%). All regression lines calculated from patient-only data pass through the normal donor data points. All CVs of the slopes of the orthogonal regression lines are well below 3%, thus fulfilling the requirements of the WHO guidelines. The mean ISI for the new CoaguChek XS PT Test is 1.01.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

An evaluation of point-of-care instrument for monitoring anticoagulation level in adult cardiac patients.

INTRODUCTION Rapid point-of-care measurement of anticoagulation has become feasible with the advent of new portable devices and offers the potential for home monitoring. This study evaluates the accuracy and feasibility of such a point-of-care device, the ProTime analyser as compared with standard laboratory method (IL MCL2) for monitoring the International Normalised Ratio (INR) level in cardi...

متن کامل

European Concerted Action on Anticoagulation. A multicentre calibration study of WHO international reference preparations for thromboplastin, rabbit (RBT/90) and human (rTF/95).

A 10 centre calibration was performed after six years to determine the international sensitivity index (ISI) of rTF/95 relative to RBT/90, and to assess any international normalised ratio (INR) bias compared with the original multicentre calibration. After exclusion of one outlying centre, the follow up calibration gave a mean ISI for rTF/95 of 0.99, which although a small difference, is signif...

متن کامل

European Concerted Action on Anticoagulation (ECAA). An assessment of lyophilised plasmas for ISI calibration of CoaguChek and TAS whole blood prothrombin time monitors.

AIMS The recommended method for the international sensitivity index (ISI) calibration of whole blood point of care testing (POCT) prothrombin time (PT) systems was originally described by Tripodi et al in 1993 but is too complex and demanding. The present European Concerted Action on Anticoagulation (ECAA) study aimed to assess the reliability of simpler ISI calibration using lyophilised plasma...

متن کامل

Poor agreement among prothrombin time international normalized ratio methods: comparison of seven commercial reagents.

BACKGROUND Prothrombin time (PT) has long been the most popular test for monitoring oral anticoagulation therapy. The International Normalized Ratio (INR) was introduced to overcome the problem of marked variation in PT results among laboratories and the various recommendations for patient care. According to this principle, all reagents should be calibrated to give identical results and the sam...

متن کامل

STABLE results: warfarin home monitoring achieves excellent INR control.

OBJECTIVES Point-of-care, home international normalized ratio (INR) monitoring (patient self-testing, or PST) provides an opportunity to optimize warfarin therapy as demonstrated in randomized trials. This study sought to determine the quality of warfarin therapy as determined by time in therapeutic INR range (TTR) in patients who perform home monitoring outside of a clinical trial setting. S...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Thrombosis and haemostasis

دوره 97 5  شماره 

صفحات  -

تاریخ انتشار 2007