Thermochromic phantom for therapeutic ultrasound daily quality assurance
نویسندگان
چکیده
Background/introduction Errors in power output ranging from −100% to +210% have been reported in a multitude of physiotherapy transducers.[1] Differences in power output can arise even after careful calibration on an annual or bi-annual schedule, which can either result in harm to the patient or non-effective treatment. Therefore, easy implementation of daily quality assurance is of great importance. We propose a simple, easy to use DQA phantom which allows the user to assess the power output of the focused ultrasound transducer, and determine if it has changed significantly after calibration. The basis of this phantom is the use of a highly attenuating ultrasound absorber with a surface layer of thermochromic liquid crystals (TLC). The use of thermochromic materials as a technique for use in ultrasound phantoms has been an ongoing area of interest. Many of these techniques require complicated set-ups, with large water tanks and imaging systems to produce very accurate data.[2,3] Simpler phantoms have been developed for use in benchtop settings, but still with a large emphasis on complicated image analysis.[4] Our proposed phantom will be a product where the end-user can visually assess the size of the lesion formed as a function of power output, without resorting to complicated image analysis.
منابع مشابه
Daily quality assurance phantom for ultrasound image guided radiation therapy
A simple phantom was designed, constructed, tested, and clinically implemented for daily quality assurance (QA) of an ultrasound-image-guided radiation therapy (US-IGRT) system, the Restitu Ultrasound system (Resonant Medical, Montreal, QC). The phantom consists of a high signal echogenic background gel surrounding a low signal hypoechoic egg-shaped target. Daily QA checks involve ultrasound im...
متن کاملUltrasound Equipment Quality Assurance
Components of an Ultrasound Quality Assurance Program 257 Quality Assurance and Preventive Maintenance 257 Tissue-Mimicking Phantoms 257 Tissue Properties Represented in Phantoms 257 Typical Quality Assurance Phantom Design 257 Cautions About Phantom Desiccation 258 Basic Quality Control Tests 259 Visual Inspection 259 Transducer Choice 259 System Sensitivity 260 Photography and Gray-Scale Hard...
متن کاملDiagnostic Ultrasound Scanner Quality Assurance
Medical Physics at CancerCare Manitoba has developed its own ultrasound phantom. The following table is intended as a quick reference guide when performing the quality assurance measurements. Instructions Limits Record Items Required ultrasound contrast-detail phantom degassed water or coupling gel paper towels for clean up (optional) lab stand and clamp to hold probe Physical check: 1. probe s...
متن کاملQuality assurance for grey-scale imaging.
In 1992, the American College of Radiology (ACR) recommended a specific phantom and image analysis procedure for (x-ray) mammography (American College of Radiology 1992). This event removed uncertainty in the USA regarding quality assurance (QA) in mammography. Although the phantom and analysis procedure are probably not ideal, existence of definitive recommendations has resulted in essentially...
متن کاملPhantom-based quality assurance measurements in B-mode ultrasound
BACKGROUND Recommended phantom-based quality assurance measurements in B-mode ultrasound (US) may be tedious. For the purpose of cost-effective US quality assurance it is important to evaluate measurements that effectively reflect the quality of US scanner. PURPOSE To find out which recommended phantom-based quality assurance measurements are effective in detecting dead or weak transducer ele...
متن کامل