Bulk etch characteristics of colorless LR 115 SSNTD
نویسندگان
چکیده
The colorless LR 115 solid-state nuclear track detector (SSNTD) (from DOSIRAD) is based on cellulose nitrate and was first studied in view of its applicability in radiobiological experiments with alpha particles. In this paper, the bulk etch characteristics were studied. We first showed that the shape of the alpha-particle tracks are irregular with blurred contours under the optical microscope. This has made measurements of track diameters very difficult. The phenomenon was explained in terms of the roughness of the detectors. As a result, the common method used to determine the bulk etch rates through measurements of opening diameters of tracks is not valid. We then proposed the surface profilometry method for determination of the removed active-layer thickness during etching. The bulk etch rates with 2.5 N aqueous solution of NaOH as the etchant, with and without magnetic stirring were determined for etching temperatures of 40, 50 and 60 C. The data supported the Arrhenius type equation. The corresponding activation energies were determined. Finally, relationships were derived between the residual active-layer thickness and the infrared transmittances at different wave numbers. The infrared transmittances were found useful in revealing the active-layer thickness of the colorless LR 115 SSNTD, the correlations being described by quadratic relationships. 2007 Elsevier B.V. All rights reserved.
منابع مشابه
Fast and non-destructive determination of active-layer thickness of LR 115 SSNTD using a color commercial document scanner
Measurements using the LR 115 solid-state nuclear track detector (SSNTD) depend critically on the removed thickness of the active layer during etching, which cannot be controlled by the etching period alone. For example, the bulk etch rate depends significantly on the strength of stirring during etching. We propose here a fast, inexpensive and nondestructive method based on a color commercial d...
متن کاملSensitivity of LR 115 SSNTD in a diffusion chamber
Solid-state nuclear track detectors (SSNTDs), such as LR 115, have been commonly used in diffusion chambers for long-term measurements of radon gas concentrations. For the LR 115 SSNTD, it has been found that the active layer removed during chemical etching is significantly affected by the presence and amount of stirring, and thus cannot be controlled easily. However, the sensitivity of the LR ...
متن کاملNon-destructive measurement of active-layer thickness of LR 115 SSNTD
LR 115 is a solid-state nuclear track detector (SSNTD) based on cellulose nitrate and has been commonly used for measurements of concentrations of radon gas and/or radon progeny. These measurements depend critically on the removed thickness of the active layer during etching. However, the thickness of removed layer calculated using the etching period does not necessarily provide a su6ciently ac...
متن کاملDerivation of V function for LR 115 SSNTD from its sensitivity to 220Rn in a diffusion chamber.
The sensitivity of the LR 115 detector inside a diffusion chamber to (220)Rn gas concentration is dependent on the removed active layer thickness during chemical etching. This dependence is related to the V function for the LR 115 detector (where V is the ratio between the track etch velocity V(t) and the bulk etch velocity V(b)) and the geometry of the diffusion chamber. The present paper pres...
متن کاملFeasibility studies of colorless LR 115 SSNTD for alpha-particle radiobiological experiments
The feasibility of using the active layer of the colorless LR 115 SSNTD for alpha-particle radiobiological experiments was studied. The track revelation time on the bottom side (the side attached to the polyester base) was much longer than that on the top side (the side not attached to the polyester base) of the active layer so track formation on the top side was more desirable. In relation to ...
متن کامل