Measurement of image rotation angle in CT for radiotherapy treatment planning
نویسندگان
چکیده
The geometric accuracy of CT images is essential when they are used for planning radiation therapy. In this study, a method of quantitative testing of image rotation in CT is presented, based on the automated analysis of an image of a phantom half-filled with water. A plug-in was written for ImageJ software for automated detection of a water surface in an image and measurement of its angle relative to a horizontal line. A water phantom can be used for the evaluation of image rotation for axial mode. In helical mode the movement of the table would cause movement of the water. The difference between image rotation for axial and helical scans can be evaluated by measuring and comparing the angles of the tabletop surface for both modes. Preliminary results of measurements for three CT scanners show that image rotation does not exceed 0.5°, and is less than 0.1° for the dedicated CT simulator. It was observed that, for one CT scanner image, the rotation angle depended on tube rotation time.
منابع مشابه
Standard edge detection algorithms versus conventional auto-contouring used for a three-dimensional rigid CT-CT matching
Background: To reduce uncertainties of patient positioning, the Computerized Tomography (CT) images acquired at the treatment planning time can be compared with those images obtained during radiation dose delivery. This can be followed during dose delivery procedure as Image Guided radiotherapy (IGRT) to verify the prescribed radiation dose delivery to the target as well as to monitor ...
متن کاملInvestigating the effects of different kernels used for CT image reconstruction on dose distributions in treatment planning of kidney cancer radiotherapy
Introduction: The quality of CT images used for treatment planning of cancer patients is an important issue in accurate outlining of the tumor volume and organs at risk. Different kernels in CT scanner systems are available for improving the image quality. Applying these kernels on CT images will change the CT numbers and electron density of tissues, conse...
متن کاملA feasibility study on the use of MV-CBCT images for urgent palliative treatment planning
Introduction: The application of 3D volumetric imaging modalities in treatment planning of radiation therapy can provide more precisely define tumor localization, and computed tomography (CT) is the most common accepted method for treatment planning. Given the lack of a CT scanner stationed in all radiotherapy departments and equipping most of the medical linear accelerators wi...
متن کاملEvaluation of changes in excision cavity volume during whole breast irradiation using secondary CT image
Background: The aim of this study was to evaluate the reduction in excision cavity volume (ECV) during the whole breast radiotherapy (WBRT). The delineating of reduced cavity volume on secondary CT (computed tomography) might decrease the field size of boost plan. Materials and Methods: Twenty patients were treated having breast conservative surgery. At first, primary CT simulation (CT-1) was d...
متن کاملImage Optimization and reduction of radiation dose in CT of the paranasal sinuses
Introduction: Since demand for computed tomography (CT) examinations is growing, patients are exposed more frequently to ionizing radiation. To reduce the dose delivered to patients in each scan, CT technologists can change the image acquisition parameters. However, this reduces image quality. Present study aimed to optimize and reduce radiation dose in a CT of the paranasal si...
متن کامل