Feasibility of a novel blood noise reduction algorithm to enhance reproducibility of ultra-high-frequency intravascular ultrasound images.
نویسندگان
چکیده
BACKGROUND Ultra-high-frequency (40- to 50-MHz) intravascular ultrasound (IVUS) improves image quality compared with conventional 20- to 30-MHz IVUS. However, as the frequency of IVUS increases, high-intensity backscatter from blood components may cause visual difficulties in discrimination between the lumen and arterial wall structure. The purpose of this study was to evaluate the effect of a novel blood noise reduction algorithm (BNR) on quantitative coronary ultrasound measurements. METHODS AND RESULTS IVUS studies using a 40-MHz transducer were performed in 35 patients with coronary artery disease. A total of 620 gray-scale images (310 pairs) were processed with and without the BNR, and lumen cross-sectional area (CSA) was determined by 2 independent observers. With the BNR, the intraobserver and interobserver correlation coefficients for lumen CSA were significantly improved (0.85 to 0.99 and 0.80 to 0.98, respectively). In the 270 images (135 pairs) in which vessel wall measurements were possible, the BNR significantly improved the intraobserver and interobserver correlation coefficients for plaque plus media CSA (0.83 to 0.99 and 0.76 to 0.97, respectively), whereas no influence was observed for external elastic membrane CSA (1.00 to 1.00 and 0.99 to 0.99, respectively). CONCLUSIONS This study demonstrates the feasibility of this novel algorithm to reduce blood noise, thereby enabling accurate lumen border delineation and providing reproducible measurements of both the lumen and plaque plus media CSAs. Incorporating a digital BNR may serve as an important adjunct to ultra-high-frequency IVUS imaging for improving accurate quantitative evaluation of vessel dimensions.
منابع مشابه
A Novel Spot-Enhancement Anisotropic Diffusion Method for the Improvement of Segmentation in Two-dimensional Gel Electrophoresis Images, Based on the Watershed Transform Algorithm
Introduction Two-dimensional gel electrophoresis (2DGE) is a powerful technique in proteomics for protein separation. In this technique, spot segmentation is an essential stage, which can be challenging due to problems such as overlapping spots, streaks, artifacts and noise. Watershed transform is one of the common methods for image segmentation. Nevertheless, in 2DGE image segmentation, the no...
متن کاملAssessment of the Wavelet Transform for Noise Reduction in Simulated PET Images
Introduction: An efficient method of tomographic imaging in nuclear medicine is positron emission tomography (PET). Compared to SPECT, PET has the advantages of higher levels of sensitivity, spatial resolution and more accurate quantification. However, high noise levels in the image limit its diagnostic utility. Noise removal in nuclear medicine is traditionally based on Fourier decomposition o...
متن کاملComparison of intravascular optical frequency domain imaging versus intravascular ultrasound during balloon pulmonary angioplasty in patients with chronic thromboembolic pulmonary hypertension
OBJECTIVES The aims of this study are (1) to evaluate the safety and feasibility of using optical frequency domain imaging (OFDI) during balloon pulmonary angioplasty (BPA) procedures, (2) to assess the correlations between the vessel area (VA) and luminal area (LA) obtained by OFDI and intravascular ultrasound (IVUS), and (3) to compare inter- and intra-observer variability among measurements ...
متن کاملEdge-Preserved Smoothing Method with Special Reference to Intravascular Ultrasound Image Using Anisotropic Diffusion Filter Controlled by Weighted Separability Measure
This paper describes a novel edge-preserved smoothing method specialized for intravascular ultrasound (IVUS) image. In the proposed method, the processing of an anisotropic diffusion filter is controlled by the weighted separability of an IVUS image. The present method not only reduces a speckle noise but also enhances an edge on an IVUS image. The effectiveness of the proposed method is verifi...
متن کاملA New Shearlet Framework for Image Denoising
Traditional noise removal methods like Non-Local Means create spurious boundaries inside regular zones. Visushrink removes too many coefficients and yields recovered images that are overly smoothed. In Bayesshrink method, sharp features are preserved. However, PSNR (Peak Signal-to-Noise Ratio) is considerably low. BLS-GSM generates some discontinuous information during the course of denoising a...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Circulation
دوره 102 14 شماره
صفحات -
تاریخ انتشار 2000