Central corneal thickness, anterior chamber depth, and pupil diameter measurements using Visante OCT, Orbscan, and Pentacam.
نویسندگان
چکیده
PURPOSE To evaluate the agreement among three different optical methods in measuring anterior chamber depth (ACD), central corneal thickness (CCT), and pupil diameter. METHODS One hundred eyes of 50 healthy patients (25 men, 25 women) were enrolled in the study. Mean patient age was 25 years (range: 21 to 32 years). Exclusion criteria were history of any intraocular or corneal surgery, contact lens wear, corneal anomalies, and spherical refraction >5.00 diopters (D) or cylindrical refraction >2.00 D. All measurements were done by the same operator under mesopic light conditions and repeated using three different optical methods: Visante optical coherence tomography (OCT) (Carl Zeiss Meditec), Orbscan (Bausch & Lomb), and Pentacam (Oculus Optikgeräte GmbH). RESULTS Mean CCT as measured by Visante OCT, Orbscan, and Pentacam was 529+/-30.5 microm, 554+/-32.7 microm, and 552+/-29.3 microm, respectively. Mean ACD values were 2.94+/-0.34 mm, 2.84+/-0.33 mm, and 2.98+/-0.33 mm, respectively. Mean pupil diameter measurements were 4.87+/-1.09 mm, 4.0+/-0.67 mm, and 3.05+/-0.59 mm, respectively. The Visante OCT measured CCT thinner and Orbscan measured ACD shallower than the other two methods. All three methods measured pupil diameters significantly different. CONCLUSIONS This study found some statistically significant but clinically insignificant differences among the optical methods assessed. The differences are small and do not influence decisions for refractive surgery in clinical practice.
منابع مشابه
Comparison of corneal measurements in keratoconic eyes using rotating Scheimpflug camera and scanning-slit topography.
AIM To compare the anterior segment measurements obtained by rotating Scheimpflug camera (Pentacam) and Scanning-slit topography (Orbscan IIz) in keratoconic eyes. METHODS A total of 121 patients, 71 males (58.7%) and 50 females (41.3%) (214 eyes) with the diagnosis of keratoconus (KC) were enrolled in this study. Following diagnosis of KC by slit-lamp biomicroscopic examination, central corn...
متن کاملComparison between Pentacam HR and Orbscan II after Hyperopic Photorefractive Keratectomy
PURPOSE The aim of this study was to determine the agreement between Pentacam HR (Scheimpflug imaging, Oculus) and Orbscan II (scanning slit topography, Bausch and Lomb) in measuring corneal parameters after photorefractive keratectomy (PRK) for hyperopia. METHODS In this prospective cross-sectional study, 38 hyperopic eyes undergoing PRK were examined before refractive surgery and 8 to 10 mo...
متن کاملCentral corneal thickness measurements using Orbscan II, Visante, ultrasound, and Pentacam pachymetry after laser in situ keratomileusis for myopia.
PURPOSE To compare corneal pachymetry assessment using 4 measurement methods in eyes after laser in situ keratomileusis (LASIK) for myopia. SETTING Department of Ophthalmology & Visual Sciences, The Chinese University of Hong Kong, Hong Kong Eye Hospital, Hong Kong SAR. METHODS Fifty-two consecutive patients (103 eyes) who had LASIK for the correction of myopia had Orbscan II (Bausch & Lomb...
متن کاملComparison of three methods of measuring corneal thickness and anterior chamber depth.
PURPOSE To compare three different methods of measuring corneal thickness (CT) and anterior chamber depth (ACD). DESIGN Prospective clinical trial (Medical University of Vienna, Austria). METHODS Central CT (CCT), CT at four peripheral points, and central ACD were measured in 88 eyes of 44 healthy subjects with the Pentacam (rotating Scheimpflug camera; Oculus, Wetzlar, Germany), Orbscan I ...
متن کاملThe distribution of orbscan indices in young population
PURPOSE To determine the distribution of anterior eye biometry indices, such as keratometry pachymetry, anterior chamber depth (ACD), pupil diameter, and corneal diameter, as measured by Orbscan instrument in a young Iranian population. METHODS A cross-sectional study was conducted, and subjects were selected through multistage cluster sampling from the students of Mashhad University of Medic...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of refractive surgery
دوره 26 2 شماره
صفحات -
تاریخ انتشار 2010