A virtual sizing tool for mitral valve annuloplasty.

نویسندگان

  • Manuel K Rausch
  • Alexander M Zöllner
  • Martin Genet
  • Brian Baillargeon
  • Wolfgang Bothe
  • E Kuhl
چکیده

Functional mitral regurgitation, a backward leakage of the mitral valve, is a result of left ventricular growth and mitral annular dilatation. Its gold standard treatment is mitral annuloplasty, the surgical reduction in mitral annular area through the implantation of annuloplasty rings. Recurrent regurgitation rates may, however, be as high as 30% and more. While the degree of annular downsizing has been linked to improved long-term outcomes, too aggressive downsizing increases the risk of ring dehiscences and significantly impairs repair durability. Here, we prototype a virtual sizing tool to quantify changes in annular dimensions, surgically induced tissue strains, mitral annular stretches, and suture forces in response to mitral annuloplasty. We create a computational model of dilated cardiomyopathy onto which we virtually implant annuloplasty rings of different sizes. Our simulations confirm the common intuition that smaller rings are more invasive to the surrounding tissue, induce higher strains, and require larger suture forces than larger rings: The total suture force was 2.2 N for a 24-mm ring, 1.9 N for a 28-mm ring, and 0.8 N for a 32-mm ring. Our model predicts the highest risk of dehiscence in the septal and postero-lateral annulus where suture forces are maximal. These regions co-localize with regional peaks in myocardial strain and annular stretch. Our study illustrates the potential of realistic predictive simulations in cardiac surgery to identify areas at risk for dehiscence, guide the selection of ring size and shape, rationalize the design of smart annuloplasty rings and, ultimately, improve long-term outcomes after surgical mitral annuloplasty. Copyright © 2016 John Wiley & Sons, Ltd.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Computer-based comparison of different methods for selecting mitral annuloplasty ring size

BACKGROUND Ring sizing for mitral valve annuloplasty is conventionally done intraoperatively using specific 'sizer' instruments, which are placed onto the valve tissue. This approach is barely reproducible since different sizing strategies have been established among surgeons. The goal of this study is to virtually apply different sizing methods on the basis of pre-repair echocardiography to fi...

متن کامل

Prediction of the annuloplasty ring size in patients undergoing mitral valve repair using real-time three-dimensional transoesophageal echocardiography

AIMS We sought to investigate the additional value of real-time three-dimensional transoesophageal echocardiography (RT 3D TOE)-guided sizing for predicting annuloplasty ring size during mitral valve repair. METHODS AND RESULTS In 53 patients undergoing elective mitral valve repair, an RT 3D TOE was performed pre- and post-operatively. The digitally stored loops were imported into a software ...

متن کامل

Surgical Treatment of Rheumatic Mitral Valve Lesions in Children and Adolescents

Rheumatismal lesions of the mitral valve a!I'e rare under the age of six and surgical iruter­vention in symptomatic patients is an academic problem. Early surgical treatment has been done to prevent serious changes and unrevers­ib1e complications o.f the heart and lungs.  From April 71 until April 74, thirty-five pati­ent, ranging between 8-18 years of age, suffer­ing of <lifferents heart ,tro...

متن کامل

Personalized Computational Modeling of Mitral Valve Prolapse: Virtual Leaflet Resection

Posterior leaflet prolapse following chordal elongation or rupture is one of the primary valvular diseases in patients with degenerative mitral valves (MVs). Quadrangular resection followed by ring annuloplasty is a reliable and reproducible surgical repair technique for treatment of posterior leaflet prolapse. Virtual MV repair simulation of leaflet resection in association with patient-specif...

متن کامل

Value of transesophageal echocardiography (TEE) guidance in minimally invasive mitral valve surgery.

The role of intraoperative transesophageal echocardiography (TEE) has increased tremendously since its first use in 1979. Today intraoperative TEE is a class I indication for surgical mitral valve reconstruction for evaluation of mitral valve pathology, graduation of mitral regurgitation and detection of potential risk factors as well as post-repair assessment. Real-time three-dimensional TEE o...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • International journal for numerical methods in biomedical engineering

دوره 33 2  شماره 

صفحات  -

تاریخ انتشار 2017