A novel approach to phase-contrast velocity offset correction by in vivo high-SNR acquisitions
نویسندگان
چکیده
Background Baseline offset errors on phase-contrast velocity images can be corrected using stationary tissue, for example subtracting fitted corrections from the image (1). Although corrections are often curved over the FOV, 1st order (linear) fitting is typical. This may partly be due to low SNR of static tissue making higher-order fitting unreliable (2). Aim: To evaluate a new method acquiring additional high SNR velocity images specifically to improve offset correction.
منابع مشابه
Part Ii Velocity Offsets: Characterization in a Multi-vendor Study
Aims: Cardiovascular magnetic resonance (CMR) allows non-invasive phase contrast measurements of flow through planes transecting large vessels. However, some clinically valuable applications are highly sensitive to errors caused by small offsets of measured velocities if these are not adequately corrected, for example by the use of static tissue or static phantom correction of the offset error....
متن کاملFlow measurement by cardiovascular magnetic resonance: a multi-centre multi-vendor study of background phase offset errors that can compromise the accuracy of derived regurgitant or shunt flow measurements
AIMS Cardiovascular magnetic resonance (CMR) allows non-invasive phase contrast measurements of flow through planes transecting large vessels. However, some clinically valuable applications are highly sensitive to errors caused by small offsets of measured velocities if these are not adequately corrected, for example by the use of static tissue or static phantom correction of the offset error. ...
متن کاملSSFP and GRE phase contrast imaging using a three-echo readout.
A technique for rapid in-plane phase-contrast imaging with high signal-to-noise ratio (SNR) is described. Velocity-encoding is achieved by oscillating the readout gradient, such that each 2DFT phase-encode is acquired three times following a single RF slice-selective excitation. Three images are reconstructed, from which both flow velocity and local resonance offset are calculated. This techniq...
متن کاملA multi-center inter-manufacturer study of the temporal stability of phase-contrast velocity mapping background offset errors
BACKGROUND Phase-contrast velocity images often contain a background or baseline offset error, which adds an unknown offset to the measured velocities. For accurate flow measurements, this offset must be shown negligible or corrected. Some correction techniques depend on replicating the clinical flow acquisition using a uniform stationary phantom, in order to measure the baseline offset at the ...
متن کاملNonlinear self-calibrated phase-contrast correction in quantitative cardiac imaging
Background Residual background phase in cardiac phase-contrast (PC) imaging introduces velocity errors that bias quantitative flow measurements [1]. While the bias can be offset using static phantoms [2], improved workflow is realized if self-calibrated correction is performed by fitting the phase of static tissue from the in vivo images. However, the residual phase can be nonlinear in space an...