The self-inflating resuscitation bag delivers high oxygen concentrations when used without a reservoir: implications for neonatal resuscitation.

نویسندگان

  • Kathy L Johnston
  • Khalid Aziz
چکیده

OBJECTIVE To measure the delivered fractional oxygen concentration (F(DO(2))) from preterm-size Laerdal silicone resuscitators (PLSR) without a reservoir. BACKGROUND The North American Neonatal Resuscitation Program manual states that self-inflating bags without a reservoir deliver approximately 40% oxygen, differing from the PLSR manufacturer's specifications. METHODS A neonatal test lung was manually ventilated using PLSRs without a reservoir. A 50 psi 100% oxygen source and an oxygen flow meter were used to provide desired oxygen inlet flows. F(DO(2)) was measured using 3 different PLSRs after 4 min of manual ventilation of a neonatal test lung, at differing inspired tidal volumes (5 mL or 20 mL), respiratory rates (40 breaths/min or 60 breaths/min), and oxygen inlet flows (1 to 4, 5, and 10 L/min). RESULTS In all tests using 5 or 10 L/min, exceeded 0.95. The lowest F(DO(2)) was 0.59, at 1 L/min. CONCLUSIONS The F(DO(2)) measured during this study did not differ from PLSR specifications. The F(DO(2)) did, however, differ from information contained in the North American Neonatal Resuscitation Program manual regarding use of a self-inflating bag without a reservoir. Care should be taken when selecting a self-inflating resuscitation device to provide blended air and oxygen, as high concentrations of oxygen may be delivered by these devices even when the reservoir is removed. American and Canadian recommendations for the provision of supplemental oxygen with self-inflating bags require reevaluation.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Oxygen outflow delivered by manually operated self-inflating resuscitation bags in patients breathing spontaneously.

OBJECTIVE To determine the oxygen outflow delivered by seven different models of manually operated self-inflating resuscitation bags (with and without an oxygen reservoir connected), which were tested using different oxygen supply rates without manipulating the bag, by simulating their use in patients breathing spontaneously. METHODS The oxygen outflow was measured using a wall oxygen flow me...

متن کامل

Innovating Neonatal Resuscitation Bags for Ease of Use in Diverse Locations.

The Chinese emperor Huang-Ti (2698–2599 BCE) noted that newborn death from ventilatory failure was common in premature infants.1 Galen (129–199 CE) inflated the lungs of deceased animals with a bellows and concluded that the air movement causes the chest to rise.1,2 Between 1850 and 1950, many techniques were used to resuscitate newborns, including swinging the infant upside down,1-3 rhythmic t...

متن کامل

A survey of delivery room resuscitation practices in the United States.

OBJECTIVE To determine current resuscitation practices of neonatologists in the United States. METHODS A 15-question survey was developed and mailed to neonatal directors in May 2004. RESULTS Of the total of 797 surveys mailed, 84 were returned undeliverable or unanswered and 450 were returned completed (63% response rate). Respondents were mainly (70%) from level III NICUs. Most programs r...

متن کامل

Alterations in peak inspiratory pressure and tidal volume delivered by manually operated self-inflating resuscitation bags as a function of the oxygen supply rate.

OBJECTIVE To assess possible alterations in the tidal volume and peak inspiratory pressure delivered by seven models of manually operated self-inflating resuscitation bags as a function of the oxygen supply rate used. METHODS The resuscitation bags tested were the following: Oxigel, models A and B; CE Reanimadores; ProtecSolutions; Missouri; Axmed; and Narcosul. For the measurements, a wall o...

متن کامل

Oxygen delivery using a neonatal self-inflating resuscitation bag: effect of oxygen flow.

OBJECTIVE We evaluated the effect of oxygen (O₂) flow rate on the corresponding delivered fraction of oxygen (FiO₂) during positive pressure ventilation (PPV) when using a neonatal self-inflating bag (SIB). METHODS Fifteen health care professionals administered PPV at a respiratory rate of 40 to 60 breaths per minute and at peak inspiratory pressures of 25 and 35 cm H₂O to a manikin by using ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Respiratory care

دوره 54 12  شماره 

صفحات  -

تاریخ انتشار 2009