2-year follow-up report on micromotion of a short tibia stem
نویسندگان
چکیده
BACKGROUND AND PURPOSE A shortened tibial stem could influence the early prosthetic fixation. We therefore compared the short stem to the standard-length stem using radiostereometric analysis (RSA) as primary outcome measure. PATIENTS AND METHODS 60 patients were randomized to receive a cemented Triathlon total knee arthroplasty (TKA) with a tibial tray of either standard or short stem length. The patients were blinded regarding treatment allocation. The micromotion of the tibial component was measured by RSA postoperatively, at 3 months, and after 1 and 2 years; clinical outcome was measured with the American Knee Society score (AKSS) and the knee osteoarthritis and injury outcome score (KOOS). RESULTS The maximum total point motion (MTPM) for the standard stem was 0.36 (SD 0.16) mm at 3 months, 0.51 (SD 0.27) mm at 1 year, and 0.54 (SD 0.28) mm at 2 years. For the short stem, it was 0.42 (0.24) mm, 0.59 (0.43) mm, and 0.61 (0.39) mm. 4 short-stemmed components and 2 standard-stemmed components had more than 0.2 mm of migration between the first- and second-year follow-up, and were classified as continuously migrating. INTERPRETATION The short-stemmed cemented tibial prosthesis showed an early prosthetic migratory pattern similar to that of the standard-stemmed cemented Triathlon knee prosthesis. The number of continuously migrating tibial plates in each group is predictive of a lower revision rate than 5% at 10 years.
منابع مشابه
Continued stabilization of Triathlon cemented TKA
Background and purpose - There is a general call for phased introduction of new implants, and one step in the introduction is an early evaluation of micromotion. We compared the micromotion in the Triathlon and its predecessor, the Duracon total knee prosthesis, concentrating especially on continuous migration over 5 years of follow-up. Patients and methods - 60 patients were randomized to rece...
متن کاملFactors associated with excessive migration in bone impaction hip revision surgery: a radiostereometric analysis study.
A prospective radiostereometric analysis (RSA) study of 18 patients with cemented revision hip surgery and impaction grafting with an Exeter stem was done with a follow-up of 2 years for all patients. All factors that could influence migration (ie, micromotion) of the stem were analyzed with a repeated measurements analysis of variance. Two groups could be identified: a stable group and a conti...
متن کاملارزیابی عملکرد بیماران پس از آرتروپلاستی هیپ با پروتز Short-stem و Standard-stem : کارآزمایی بالینی تصادفی
Background: Short-stem prosthesis for total hip arthroplasty (THA) have been designed to overcome the weakness of standard-stem prostheses and improve surgical outcomes. The aim of this study was to compare short-stem with standard-stem prosthesis outcomes. Methods: This study was performed as a randomized clinical trial. Subjects were selected among patients referred to Sina University Hospit...
متن کاملMelorheostosis in proximal phalanx and metacarpal bone: a short-mid-term follow up
Melorheostosis is a rare sclerosing hyperostosis that affects the appendicular skeleton more than the axial skeleton. This disorder also affects the cutaneous and soft tissues and leads to the formation of fibrosis and contracture. Due to the unique features of melorheostosis (dripping candle wax appearance), plain radiographs are often sufficient for diagnosis. Herein we report a case of melor...
متن کاملBacille Calmette-Guérin osteomyelitis
Background: Tuberculosis is an essential problem for healthcare systems especially in developing countries . All newborns are given Bacillus Calmette-Guérin (BCG) vaccine in Iran which is prepared from live bovine tuberculosis bacillus, and is given to protect against tuberculosis. Osteomyelitis secondary to BCG vaccination is rare and usually involves epiphysis of long tubular bones. Methods...
متن کامل