Robot-assisted stereotactic laser ablation in medically intractable epilepsy: operative technique.

نویسندگان

  • Jorge Gonzalez-Martinez
  • Sumeet Vadera
  • Jeffrey Mullin
  • Rei Enatsu
  • Andreas V Alexopoulos
  • Ravish Patwardhan
  • William Bingaman
  • Imad Najm
چکیده

BACKGROUND Stereotactic laser ablation offers an advantage over open surgical procedures for treatment of epileptic foci, tumors, and other brain pathology. Robot-assisted stereotactic laser ablation could offer an accurate, efficient, minimally invasive, and safe method for placement of an ablation catheter into the target. OBJECTIVE To determine the feasibility of placement of a stereotactic laser ablation catheter into a brain lesion with the use of robotic assistance, via a safe, accurate, efficient, and minimally invasive manner. METHODS A laser ablation catheter (Visualase, Inc) was placed by using robotic guidance (ROSA, Medtech Surgical, Inc) under general anesthesia into a localized epileptogenic periventricular heterotopic lesion in a 19-year-old woman with 10-year refractory focal seizure history. The laser applicator (1.65 mm diameter) position was confirmed by using magnetic resonance imaging (MRI). Ablation using the Visualase system was performed under multiplanar imaging with real-time thermal imaging and treatment estimates in each plane. A postablation MRI sequence (T1 postgadolinium contrast injection) was used to immediately confirm the ablation. RESULTS MRI showed accurate skin entry point and trajectory, with the applicator advanced to the lesion's distal boundary. Ablation was accomplished in less than 3 minutes of heating. The overall procedure, from time of skin incision to end of last ablation, was approximately 90 minutes. After confirmation of proper lesioning by using a T1 contrast-enhanced MRI, the applicator was removed, and the incision was closed using a single stitch. No hemorrhage or other untoward complication was visualized. The patient awoke without any complication, was observed overnight after admitting to a regular floor bed, and was discharged to home the following day. CONCLUSION This technique, using a combination of Visualase laser ablation, ROSA robot, and intraoperative MRI, facilitated a safe, efficacious, efficient, and minimally invasive approach that could be used for placement of 1 or multiple electrodes in the future.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Stereotactic Laser Ablation for Medically Intractable Epilepsy: The Next Generation of Minimally Invasive Epilepsy Surgery

Epilepsy is a common, disabling illness that is refractory to medical treatment in approximately one-third of patients, particularly among those with mesial temporal lobe epilepsy. While standard open mesial temporal resection is effective, achieving seizure freedom in most patients, efforts to develop safer, minimally invasive techniques have been underway for over half a century. Stereotactic...

متن کامل

Robot Assisted Stereotactic Laser Ablation for a Radiosurgery Resistant Hypothalamic Hamartoma

Hypothalamic hamartomas (HH) are benign tumors that can cause significant morbidity in adults as a cause of epilepsy, particularly gelastic seizures. Open and endoscopic resections of HH offer good seizure control but have high rates of morbidity and are technically challenging. Stereotactic radiosurgery has been an alternative treatment; however, it results in comparably poor seizure control. ...

متن کامل

Stereotactic laser ablation of the splenium for intractable epilepsy☆

Partial or complete corpus callosotomies have been applied, traditionally via open surgical or radiosurgical approaches, for the treatment of epilepsy in patients with multifocal tonic, atonic, or myoclonic seizures. Minimally invasive methods, such as MRI-guided laser interstitial thermal ablation (MTLA), are being employed to functionally remove or ablate seizure foci in the treatment of epil...

متن کامل

Technique, Results, and Complications Related to Robot-Assisted Stereoelectroencephalography.

BACKGROUND Robot-assisted stereoelectroencephalography (SEEG) may represent a simplified, precise, and safe alternative to the more traditional SEEG techniques. OBJECTIVE To report our clinical experience with robotic SEEG implantation and to define its utility in the management of patients with medically refractory epilepsy. METHODS The prospective observational analyses included all patie...

متن کامل

Laser ablation as treatment strategy for medically refractory dominant insular epilepsy: therapeutic and functional considerations.

Since its introduction to neurosurgery in 2008, laser ablative techniques have been largely confined to the management of unresectable tumors. Application of this technology for the management of focal epilepsy in the adult population has not been fully explored. Given that nearly 1,000,000 Americans live with medically refractory epilepsy and current surgical techniques only address a fraction...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Neurosurgery

دوره 10 Suppl 2  شماره 

صفحات  -

تاریخ انتشار 2014