Performance of Lead Integrity Alert to assist in the clinical diagnosis of implantable cardioverter defibrillator lead failures: analysis of different implantable cardioverter defibrillator leads.
نویسندگان
چکیده
BACKGROUND The Lead Integrity Alert (LIA) was developed for Medtronic implantable cardioverter defibrillators to reduce inappropriate shocks for rapid oversensing caused by conductor fractures and reported for Medtronic Fidelis conductor fractures. The goal of this study was to compare the performance of LIA with conventional impedance monitoring for identifying lead system events (LSEs) and lead failures (LFs) in lead families that differ from Fidelis. METHODS AND RESULTS We analyzed data from 12 793 LIA enabled implantable cardioverter-defibrillator and lead combinations including 6123 St. Jude Riata or Durata, 5114 Boston Scientific Endotak, and 1556 Fidelis combinations followed in the CareLink remote monitoring network for LSEs and LFs. Each alert was adjudicated based on implantable cardioverter-defibrillator stored electrograms/diagnostics and clinical data as an LSE or non-lead system event by 2 physicians after reviewing the electrograms and clinical data. During 13 562 patient-years of LIA follow-up, there were 179 adjudicated alerts, of which 84 were LSEs (including 65 LFs) and 95 were non-lead system events. LIA identified >66% more LSE and >67% more LF compared with conventional impedance monitoring and did not differ by lead family for LSE (P=0.573) or LF (P=0.332). Isolated spikes on electrogram were associated more often with LF in St. Jude leads (71%) compared with Endotak (9%; P<0.001) and Fidelis leads (11%; P<0.001). The non-lead system event detection rate was different among lead families (P<0.001) ranging from 1 in every 78.5 years (Endotak), 228.9 years (St. Jude leads), and 627.6 years (Fidelis). CONCLUSIONS LIA markedly increased the detection rate of LSE compared with conventional impedance monitoring.
منابع مشابه
THE EFFECT OF NURSE-PEER-LED SUPPORT INTERVENTION ON QUALITY OF LIFE IN PATIENTS WITH AN IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR: A RANDOMIZED CLINICAL TRIAL
Background & Aims: Nowadays, the quality of life and how coping and treatment of patients with an implantable cardiac defibrillator has led to concern and attempts to reduce or eliminate the problems of these patients in the people who provide health care services. This study aimed to determine the effect of nurse-peer-led support intervention on quality of life in patients with an implantable ...
متن کاملImplantable cardioverter-defibrillator in a patient with dextrocardia situs inversus
Background: Dextrocardia is a congenital anomaly, which may have coexistent coronary artery disease (CAD), arrhythmias and conventional indications for device therapy. However, the implantation of transvenous leads can be technically challenging and the approach needs to be tailored to the patient's individual anatomy. Case presentation: A 54-year-old male with dextrocardia situs inversus and i...
متن کاملزندگی با دفیبریلاتور قابل کاشت: یک مطالعه پدیدهشناسی
Background & Aim: Implantable Cardioverter Defibrillator is the most effective treatment for cardiac dysrhythmias. Having such life-saving device is the unique experience with important changes in the patient’ life. Understanding of these experiences can lead to success in treatment and follow- up.This study aimed to understand the experiences of patients with Implantable Cardioverter Defibri...
متن کاملIs correction for metallic artefacts mandatory in cardiac SPECT/CT imaging in the presence of pacemaker and implantable cardioverter defibrillator leads?
Introduction: Metallic artifacts due to pacemaker/ implantable cardioverter defibrillator (ICD) leads in CT images can produce artifactual uptake in cardiac SPECT/CT images. The aim of this study was to determine the influence of the metallic artifacts due to pacemaker and ICD leads on myocardial SPECT/CT imaging. Methods: The study included 9 patients wh...
متن کاملMyopotential oversensing notified by Lead Integrity Alert in a patient with implantable cardioverter defibrillator with a dedicated bipolar epicardial sensing lead
Although myopotential oversensing by a dedicated bipolar lead is rare, an epicardial lead on a dilated ventricle might contribute to its sensitivity. Myopotential oversensing was notified by the Lead Integrity Alert in this case. We should be aware of this possibility for the management of such patients.
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Circulation. Arrhythmia and electrophysiology
دوره 6 6 شماره
صفحات -
تاریخ انتشار 2013