Comparison of Cervical Range-of-Motion Restriction and Craniofacial Tissue-Interface Pressure With 2 Adjustable and 2 Standard Cervical Collars.

نویسندگان

  • Ann N Tescher
  • Aaron B Rindflesch
  • James W Youdas
  • Ross W Terman
  • Therese M Jacobson
  • Lisa L Douglas
  • Anne G Miers
  • Christine M Austin
  • Adriana M Delgado
  • Savannah M Zins
  • Brian D Lahr
  • Mark A Pichelmann
  • Stephanie F Heller
  • Paul M Huddleston
چکیده

STUDY DESIGN Randomized controlled trial. OBJECTIVE The aim of the study was to compare and contrast the restrictiveness and tissue-interface pressure (TIP) characteristics of 2 standard and 2 adjustable cervical collars. SUMMARY OF BACKGROUND DATA This study compared the restrictiveness and TIP of 4 commercially available cervical collars (2 standard and 2 adjustable). Adjustable collars offer potential advantages of individualized fit for patients and decreased inventory for institutions. The overall goal was to determine whether the adjustable collars provided the same benefits of cervical range-of-motion (CROM) restriction as the standard collars without increasing TIP and risk of pressure-related complications. METHODS A total of 48 adult volunteer subjects (24 men and 24 women) were fitted with 4 collars (Aspen, Aspen Vista, Miami J, and Miami J Advanced) in random order. Data collection included assessment of CROM restrictiveness and measurement of TIP on the mandible and occiput in upright and supine positions. The experimental, repeated measures design stratified the sample by body mass index (BMI) and sex. RESULTS All collars restricted CROM as compared with no collar (P ≤ 0.001 each). Aspen was more restrictive than Aspen Vista and Miami J in 4 movement planes (P ≤ 0.003 each), but not significantly different from Miami J Advanced. The Miami J standard collar was associated with significantly lower peak TIPs on all sites and in all positions compared with Aspen (P ≤ 0.001), Miami J Advanced (P < 0.001), and Aspen Vista (P = 0.01 for mandible site and upright position, P < 0.001 for remaining sites and positions). Increased peak TIP correlated with high BMI across all collar types, but was significantly lower for the Miami J collar than the Aspen collar. CONCLUSION All collars, compared with no collar, significantly restricted CROM. Although the collar-to-collar comparisons were statistically significant, the differences may have little clinical significance in the acutely injured trauma patient. The Miami J standard collar had the lowest overall TIP in both sites and positions. Ongoing effort should be devoted to staff education in proper sizing and fit, particularly for patients with high BMI.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Effect of cervical collars on intracranial pressure in patients with head neurotrauma

Trauma patients are at high risk of cervical and associated injuries. The standard protocol includes immobilization of the cervical spine in a polytrauma as it will help to prevent spinal injuries in the prehospital settings. Hard cervical collar (or alternatives) is routinely used until the cervical spine clearance is obtained . In patients with traumatic brain injury avoiding any kind of mane...

متن کامل

Biomechanical analysis of cervical orthoses in flexion and extension: a comparison of cervical collars and cervical thoracic orthoses.

The analysis of current cervical collars (Aspen and Miami J collars) and cervical thoracic orthoses (CTOs) (Aspen 2-post and Aspen 4-post CTOs) in reducing cervical intervertebral and gross range of motion in flexion and extension was performed using 20 normal volunteer subjects. The gross sagittal motion of the head was measured relative to the horizon with the use of an optoelectronic motion ...

متن کامل

Immediate Effect of Kinesio-Taping on Cervical Lateral Flexion Range of Motion in Subjects With Myofascial Trigger Point in Upper Trapezius Muscle

Purpose: The purpose of this study was to investigate the effect of kinesio-taping on the range of motion of cervical lateral flexion point in subjects with trigger point in upper trapezius muscle. Methods: A total of 32 subjects with myofascial trigger point in upper trapezius participated in this randomized controlled trial (RCT). The subjects were randomly assigned to two groups: Kinesio-...

متن کامل

The Reliability and Concurrent Validity of Digital Inclinometer, Smartphone Applications, and the Cervical Range of Motion Device for Measuring the Cervical Range of Motion

Objectives: Changes in the Range of Motion (ROM) are essential criteria in determining the severity of spinal disorders and could be effective in predicting pain progression. Instruments to measure the ROM are costly and unavailable in most therapy settings. While there is a tendency in therapists to use their smartphones instead, there is no report to measure the suitability of smartphones to ...

متن کامل

The measurement of tissue interface pressures and changes in jugular venous parameters associated with cervical immobilisation devices: a systematic review

Cervical immobilisation is commonly applied following trauma, particularly blunt head injury, but current methods of immobilisation are associated with significant complications. Semi-rigid disposable cervical collars are known to cause pressure ulcers, and impede effective airway management. These collars may also exacerbate a head injury by increasing intracranial pressure as a result of exte...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Spine

دوره 41 6  شماره 

صفحات  -

تاریخ انتشار 2016