The impact of CT scan energy on range calculation in proton therapy planning

نویسندگان

  • Kevin K. Grantham
  • Hua Li
  • Tianyu Zhao
  • Eric E. Klein
چکیده

The purpose of this study was to investigate the impact of tube potential (kVp) on the CT number (HU) to proton stopping power ratio (PSPR) conversion. The range and dosimetric change introduced by a mismatch in kVp used for the CT scan and the HU to PSPR table, based on a specific kVp, used to calculate dose are analyzed. Three HU to PSPR curves, corresponding to three kVp settings on the CT scanner, were created. A treatment plan was created for a single beam in a water phantom passing through a wedge-shaped bone heterogeneity. The dose was recalculated by changing only the HU to PSPR table used in the dose calculation. The change in the position of the distal 90% isodose line was recorded as a function of heterogeneity thickness along the beam path. The dosimetric impact of a mismatch in kVp between the CT and the HU to PSPR table was investigated by repeating this procedure for five clinical plans comparing DVH data and dose difference distributions. The HU to PSPR tables diverge for CT numbers greater than 200 HU. In the phantom plan, the divergence of the tables resulted in a difference in range of 1.6 mm per cm of bone in the beam path, for the HU used. For the clinical plans, the dosimetric effect of a kVp mismatch depends on the amount of bone in the beam path and the proximity of OARs to the distal range of the planned beams. A mismatch in kVp between the CT and the HU to PSPR table can introduce inaccuracy in the proton beam range. For dense bone, the measured range difference was approximately 1.6 mm per cm of bone along the beam path. However, the clinical cases analyzed showed a range change of 1 mm or less. Caution is merited when such a mismatch may occur.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Impact of Various Beam Parameters on Lateral Scattering in Proton and Carbon-ion Therapy

Background: In radiation therapy with ion beams, lateral distributions of absorbed dose in the tissue are important. Heavy ion therapy, such as carbon-ion therapy, is a novel technique of high-precision external radiotherapy which has advantages over proton therapy in terms of dose locality and biological effectiveness.Methods: In this study, we used Monte Carlo method-based Geant4 toolkit to s...

متن کامل

Calculation of Neutron Dose Ratio of Heart, Lung and Liver due to breast cancer Proton Therapy using MCNPX code

Introduction: The proton beam produced in particle accelerators has an appropriate therapeutic potential. In this research, proton therapy of breast cancer is simulated using the MCNPX code in a MIRD phantom, also the contribution of scattered neutron dose during the proton therapy were calculated for the Heart, Lung and Liver.   Materials and Methods: For si...

متن کامل

Monte Carlo calculation of proton ranges in water phantom for therapeutic energies

Introduction: One crucial point when calculating the distribution of doses with ions is the uncertainty of the Bragg peak. The proton ranges in determined geometries like homogeneous phantoms and detector geometries can be calculated with a number of various parameterization models. Several different parameterizations of the range-energy relationship exist, with different level...

متن کامل

Impact of range straggling and multiple scattering on proton therapy of brain, using a slab head phantom

Background: The advantages of proton beam in radiation therapy- like small lateral scattering as well as absence of exit dose tail in the organs which are after the tumor- make it capable of delivering more treatment doses to the target and much lesser to the critical tissues near it. Materials and Methods: In this study, the Monte Carlo MCNPX code has been used to simulate a slab head phantom ...

متن کامل

Evaluation of Effect of Different Computed Tomography Scanning Protocols on Hounsfield Unit and Its Impact on Dose Calculation by Treatment Planning System

Introduction: In radiotherapy treatment planning system (TPS), basic input is the data from computed tomography (CT) scan, which takes into account the effect of inhomogeneities in dose calculations. Measurement of CT numbers may be affected by scanner-specific parameters. Therefore, it is important to verify the effect of different CT scanning protocols on Hounsfield unit (HU) and its impact o...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 16  شماره 

صفحات  -

تاریخ انتشار 2015