OBJECTIVE To systematically evaluate the effects of laser beam size, power, and pulse duration of 1 to 100 milliseconds on the characteristics of ophthalmoscopically visible retinal coagulation lesions. METHODS A 532-nm Nd:YAG laser was used to irradiate 36 retinas in Dutch Belt rabbits with retinal beam sizes of 66, 132, and 330 mum. Lesions were clinically graded 1 minute after placement, their size measured by digital imaging, and their depth assessed histologically at different time points. RESULTS Retinal lesion size increased linearly with laser power and logarithmically with pulse duration. The width of the therapeutic window, defined by the ratio of the threshold power for producing a rupture to that of a mild coagulation, decreased with decreasing pulse durations. For 132- and 330-mum retinal beam sizes, the therapeutic window declined from 3.9 to 3.0 and 5.4 to 3.7, respectively, as pulse duration decreased from 100 to 20 ms. At pulse durations of 1 millisecond, the therapeutic window decreased to unity, at which point rupture and a mild lesion were equally likely to occur. CONCLUSIONS At shorter pulse durations, the width and axial extent of the retinal lesions are smaller and less dependent on variations in laser power than at longer durations. The width of the therapeutic window, a measure of relative safety, increases with the beam size. CLINICAL RELEVANCE Pulse durations of approximately 20 milliseconds represent an optimal compromise between the favorable impact of speed, higher spatial localization, and reduced collateral damage on one hand, and sufficient width of the therapeutic window (> 3) on the other.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Effects of Laser Physical Parameters on Lesion Size in Retinal Photocoagulation Surgery: Clinical OCT and Experimental Study

Introduction: The aim of the present study was to determine burn intensity in retinal laser photocoagulation based on laser parameters; wavelength, power, beam size and pulse duration, using Optical Coherence Tomography (OCT), fundus camera, physical eye model and computer simulation in a clinical study.Materials and Methods: Participants were 10 adult patients between 50-80 years with prolifer...

متن کامل

Effects of Laser Physical Parameters on Lesion Size in Retinal Photocoagulation Surgery: Clinical OCT and Experimental Study

Introduction The aim of the present study was to determine burn intensity in retinal laser photocoagulation based on laser parameters; wavelength, power, beam size and pulse duration, using Optical Coherence Tomography (OCT), fundus camera, physical eye model and computer simulation in a clinical study. Materials and Methods Participants were 10 adult patients between 50-80 years with prolife...

متن کامل

THERAPEUTIC EFFECTS OF LASER PHOTOCOAGULATION AND VITRECTOMY IN EALES\'DISEASE

Eales' disease is an idiopathic obliterative retinal vasculopathy that is treated with laser photocoagulation, and/or vitrectomy. In order to determine visual outcomes and regression of retinal neovascularization following scattered peripheral retinal photocoagulation, and/or vitrectomy in cases with Eales' disease, we did a retrospective study of the records of these patients treated in La...

متن کامل

ynamics of retinal photocoagulation and rupture

aniel Palanker tanford University epartment of Ophthalmology and ansen Experimental Physics Laboratory 52 Lomita Mall, Room 135 tanford, California 94305 Abstract. In laser retinal photocoagulation, short 20 ms pulses have been found to reduce thermal damage to the inner retina, decrease treatment time, and minimize pain. However, the safe therapeutic window defined as the ratio of power for pr...

متن کامل

The impact of pulse duration and burn grade on size of retinal photocoagulation lesion: implications for pattern density.

PURPOSE Shorter pulses used in pattern scanning photocoagulation (10-20 milliseconds [ms]) tend to produce lighter and smaller lesions than the Early Treatment Diabetic Retinopathy Study standard 100-ms exposures. Smaller lesions result in fewer complications but may potentially reduce clinical efficacy. It is worthwhile to reevaluate existing standards for the number and size of lesions needed...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Archives of ophthalmology

دوره 126 1  شماره 

صفحات  -

تاریخ انتشار 2008