Long-term temporary pacing with an active fixation lead.

نویسندگان

  • Aleksander Maciąg
  • Paweł Syska
  • Artur Oręziak
  • Andrzej Przybylski
  • Beata Broy
  • Piotr Kołsut
  • Dariusz Zając
  • Maria Bilińska
  • Maciej Sterliński
  • Hanna Szwed
چکیده

BACKGROUND Ensuring a haemodynamically effective cardiac rhythm is a challenge in patients waiting for pacemaker reimplantation after transcutaneous lead extraction due to an infection of the implanted system. AIM The authors report a retrospective analysis of temporary pacing with an active fixation lead (AFTP) connected to an externalised pacemaker in patients after transvenous lead extraction (TLE) due to an infection. METHODS AFTP was used in 34 patients (12 women) aged from 38 to 88 years (mean 67.5 years). This represented 24.5% of the population of patients undergoing TLE due to infective indications. In 32 cases, the indication for temporary pacing was atrioventricular block, and in 2 patients sick sinus syndrome. The lead was implanted via the internal jugular vein puncture into the right ventricle in 33 cases and into the right atrium in 1 case. Leads were secured to the skin and attached to externalized pacemakers. RESULTS AFTP was used for 4 to 26 days (average 14.5 days). Re-implantation was performed in 29 patients (85.3% of the study group). There was no early infection recurrence. Three patients died during AFTP (8.8% of the study group), including two due to septic shock, and a cardiac arrest due to pulseless electrical activity in another patient. CONCLUSIONS Temporary pacing with an active fixation lead is an effective and safe method to ensure a hemodynamically stable heart rhythm for a period ranging from a few to several days after the surgery in patients after transcutaneous lead extraction due to infective indications.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A long-term, prospective, cohort study on the performance of right ventricular pacing leads: comparison of active-fixation with passive-fixation leads

Active-fixation pacing leads allow the use of selective pacing sites. We evaluated their long-term performance versus passive-fixation leads in 199 newly implanted patients (n = 100 active and n = 99 passive). Postoperative pacing thresholds in the active group were higher than in the passive group (0.85 ± 0.31 V vs. 0.53 ± 0.21 V at baseline, P < 0.001). The active thresholds fell to 0.72 ± 0....

متن کامل

Morphology Of Current Of Injury Does Not Predict Long Term Active Fixation ICD Lead Performance

BACKGROUND Currents of injury (COI) have been associated with improved lead performance during perioperative measurements in pacemaker and ICD implants. Their relevance on long term lead stability remains unclear. METHODS Unipolar signals were recorded immediately after active fixation ICD lead positioning, blinded to the implanting surgeon. Signals were assigned to prespecified COI types by ...

متن کامل

Targeted left ventricular endocardial pacing using a steerable introducing guide catheter and active fixation pacing lead.

AIMS Cardiac resynchronization therapy via the coronary sinus (CS) is not always possible. Left ventricular (LV) endocardial lead placement is a potential alternative. The purpose of this study was to assess the feasibility of endocardial LV pacing using a steerable lead introducer and active fixation polyurethane lead. METHODS AND RESULTS Endocardial LV lead placement was attempted in nine p...

متن کامل

Long-term performance of the Medtronic Sprint Fidelis lead: a matter of lead type?

AIMS In 2007, the Medtronic™ Sprint Fidelis(®) lead was withdrawn from the market due to elevated failure rates. Since then, several studies were published with failure rates between 1.3 and 3.75%/year. However, they included a very high percentage of active fixation leads. Data in a population with passive leads are missing. METHODS AND RESULTS All 166 patients who received a Fidelis lead be...

متن کامل

Long-term reliability of sweet-tip type screw-in leads

BACKGROUND Active fixation leads have provided stable atrial and ventricular pacing; however, long-term follow-up data have not been satisfactory. The purpose of this study was to investigate the long-term reliability of active fixation leads and their electrical characteristic stability. METHODS A total of 1196 pacing leads were implanted in 830 patients consecutively between 2002 and 2013. ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Kardiologia polska

دوره 73 12  شماره 

صفحات  -

تاریخ انتشار 2015