Rational design of potentiometric trace level ion sensors. A Ag+-selective electrode with a 100 ppt detection limit.

نویسندگان

  • Alan Ceresa
  • Aleksandar Radu
  • Shane Peper
  • Eric Bakker
  • Ernö Pretsch
چکیده

Submicromolar to picomolar lower detection limits have recently been obtained with various polymer membrane ion-selective electrodes by minimizing biases due to ion fluxes through the membrane. For the best performance, the compositions of the membrane and inner solution should be optimized for each application. Given the number of parameters to be adjusted, it has been difficult to find the best parameters for a target sample. In this paper, a much simplified and more practical steady-state model of zero-current ion fluxes is derived, which is based on measurable parameters. The model allows one to predict achievable lower detection limits for a membrane with given selectivities. It can also be used to predict the optimal composition of the inner filling solution for the measurement of samples with a known, typical ionic background. Selectivity coefficients of monovalent and divalent analyte ions required for desired detection limits in drinking water are calculated. As an application of the proposed general recipe, a silver-selective electrode is developed on the basis of the ionophore O,O''-bis[2-(methylthio)ethyl]-tert-butylcalix[4]arene. With the predicted optimal composition of the inner electrolyte, its lower detection limit is found to be 10(-9) M or 100 ppt Ag+ with an ionic background of 10(-5) M LiNO3, which is very close to the expected value.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Al(III) Coated Wire Selective Electrode Based on 5-Bromo(salicylidene-2-aminothiophenol) Schiff Base as a New Ionophore

A new Al(III) selective membrane based on 5-bromo(salicylidene-2-amino thiophenol) as a neutral carrier in a poly(vinyl chloride) (PVC) matrix is coated on platinum wire to determine trace amounts of Al(III) ion in various samples. The sensor exhibits a good response for Al(III) ion over a wide concentration range of 5.20 × 10-7 to 1.00 × 10-2 M  with Nernstian slope of 19...

متن کامل

Determination of trace amounts of Ag (I) in waste water samples by a novel potentiometric sensor based on perphenazine as an ionophore

Background and Objective: Silver is a toxic heavy metal that is used in various industries and has adverse effects on both human health and the environment. In this respect its determination with sensitive and economic analytical methods is of great importance. Materials and Methods: In this research, a novel ion selective electrode based on perphenazine as an ionophore was developed for deter...

متن کامل

Iodide selective membrane electrode based on copper (Π)-bis-Nphenilsalicyldenaminato complex

An iodide ion-selective PVC membrane sensor based on copper (Π)-bis-Nphenilsalicyldenaminatocomplex as a novel sensing material is successfully developed. Theelectrode showed a good selectivity for iodide ion with respect to common inorganic anions. Thesensor exhibited a good linear response with slope of _58.6 ± 0.4 mV per decade over theconcentration range of 5.0 × 10-6 to 1.0 × 10-1 mol L-1,...

متن کامل

Ion-Selective Carbon Paste Electrode Based on 2-Amino-N-benzthioazolyl Benzamide (ABTB) for Determination of Copper (II) by Potentiometric Method

2-Amino-N-benzthioazolyl benzamide (ABTB) was synthesized, characterized and used for thefabrication of a potentiometric sensor for Cu2+ metal ions. The electrode exhibits linear response to Cu(II) over a wide concentration range (4.79×10 -8 – 1.85×10 -1 M) with Nernstian slope of 30 ± 1.5 mVper decade. The electrode can be used in the pH range from 2 to 9. It has a fast response time of about1...

متن کامل

Uultra-trace determination of cadmium (II) in aqueous environmental samples by a new ion selective electrode based on nefazodone and 1-butyl-3-methylimidazolium hexafluorophosphate ionic liquid

Background and Objective: Cadmium is a potential environmental contaminant that causes adverse effects on the environment and the health of living organisms. Therefore, designing a simple and economic technique for its determination is very important. In this respect, a potentiometric sensor based on nefazodone as the ionophore and [BMIM]PF6 as the ionic additive were developed for the determin...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Analytical chemistry

دوره 74 16  شماره 

صفحات  -

تاریخ انتشار 2002