Comparative Evaluation of Accuracy of Three Electronic Apex Locators in Different Simulated Clinical Conditions- An invitro Study
نویسندگان
چکیده
How to Cite this Article: Taneja S, Kumar M, Sharma SS, Gogia H, et al. Comparative Evaluation of Accuracy of Three Electronic Apex Locators in Different Simulated Clinical ConditionsAn invitro Study. Ann Med Health Sci Res. 2017; 7: 190-194 This is an open access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 3.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as the author is credited and the new creations are licensed under the identical terms. Abstract Background: The establishment of appropriate working length is one of the most critical steps in endodontic therapy. There is alteration in accuracy of apex locators in presence of irrigants, pulp/blood and open apices. Hence their accuracy in these conditions needs to be evaluated. Aim: The aim of this in vitro study was to determine the accuracy in measuring the working length by using CanalPro, Apex ID and Root ZX mini Apex locators in different simulated clinical conditions. Materials and Methods: 60 freshly extracted single-rooted teeth were equally assigned to three groups according to simulated clinical conditions: Group 1; Presence of irrigant, Group 2; open apex, Group 3; presence of blood & pulp. The working length was determined with all the apex locators and was then compared with actual working length of the tooth. The difference between the length measured by electronic apex locator (EAL) and actual length (AL) was tabulated and statistically analyzed using one way ANOVA and Post hoc Bonferroni test. Results: The accuracy in working length measurement by Root ZX mini was 100%, 60% and 80%; Apex ID was 100%, 80% and 90% and CanalPro was 100%, 90% and 90% in group 1, group 2 and group 3 respectively within ± 0.5 mm of AL. ANOVA revealed that comparison of the mean difference between length measured by all EAL’s and AL was statistically insignificant in group I. In group II and III, CanalPro showed the least difference between AL and length measured by it though it was statistically insignificant as compared to Apex ID. Root ZX showed the maximum difference between actual length and length measured by it. Conclusion: CanalPro showed highest accuracy in all conditions with accepted accuracy percentage above 90%.
منابع مشابه
In vitro evaluation of the accuracy of two new electronic apex locators in root canal length determination
In vitro evaluation of the accuracy of two new electronic apex locators in root canal length determination Dr. Hasheminia, S.M.* - Dr. Jafari, N.** *Associate Professor of Endodontics Dept., Faculty of Dentistry, Isfahan University of Medical Sciences. ** Dentist. Abstract Background & Aim: Radiography is still the common method of measuring working length in root canal therapy. While, gag refl...
متن کاملThe effect of different needle-insertion depths on the accuracy of the Root ZX II and Root ZX mini apex locators in the presence of various irrigants
BACKGROUND AND AIM: The aim of the present study was to evaluate the effect of different needle-insertion depths on the accuracy of the Root ZX II and Root ZX mini apex locators in the presence of various irrigants. METHODS: Ninety extracted single-canal human teeth were used in nine experimental groups. The coronal one-third of the canals was preflared, and the length to the major foramen was ...
متن کاملAn Invitro Evaluation of the Accuracy of Two Electronic Apex Locators to Determine Working Length in the Presence of Various Irrigants
BACKGROUND Cleaning and shaping of root canal system requires various irrigating solutions. The question is whether the presence of irrigation solution alters working length determination using apex locators. The purpose of this study was to compare the accuracy of Root ZX and i-Root apex locator for determining working length in the presence of different irrigating solution. MATERIALS AND ME...
متن کاملEvaluation of the accuracy of two apex locators in endodontic treatment and retreatments: an ex vivo study
Introduction: Working length determination is important in successful endodontic treatment and retreatment.This study evaluated the accuracy of two electronic apex locators Root ZX and Raypex®6 (EALs) in determining the electronic working length (EWL) of the root canals in endodontic treatment and retreatment. Materials and Methods: Access cavities were prepared on forty extracted, singl...
متن کاملComparative evaluation of the accuracy of two electronic apex locators in determining the working length in teeth with simulated apical root resorption: An in vitro study
INTRODUCTION Accurate determination of working length (WL) is a critical factor for endodontic success. This is commonly achieved using an apex locator which is influenced by the presence or absence of the apical constriction. Hence, this study was done to compare the accuracy of two generations of apex locators in teeth with simulated apical root resorption. MATERIALS AND METHODS Forty maxil...